Feasibility of noninvasive hemodynamic monitoring by bioreactance for air-evacuated casualties
Authors:
- Clément DubostCDClément Dubost
- Fabien SauvetFSFabien Sauvet
- David EvansDEDavid Evans
- Sandra DusonchetSDSandra Dusonchet
- Christophe PelletierCPChristophe Pelletier
- Jean-Marie RousseauJRJean-Marie Rousseau
- Sébastien CosteSCSébastien Coste
- 1 April 2013
- journal article
- research article
- Published by Wolters Kluwer Health in Journal of Trauma and Acute Care Surgery
- Vol. 74 (4) , 1146-1150
Abstract
The purpose of this study was to assess the feasibility of hemodynamic monitoring by bioreactance during the air evacuation of war casualties. From June 2011 to January 2012, the French Air Force used a noninvasive bioreactance monitor during the aeromedical evacuation of patients with severe battle injury. Patients were continuously monitored from takeoff (baseline) until arrival. Cardiac index, quality, and viability of signal and fluctuations in values were assessed during the entire flight. Eighteen patients were included, and monitoring was possible for 16 patients. The implementation of the monitoring was found to be straightforward for both doctors and paramedics, and we observed no loss of signal or interference during the flights. Cardiac index decreased significantly during the takeoff and returned to baseline value after the 50th minute of flight. The application of noninvasive bioreactance hemodynamic monitoring is feasible and pertinent in the specific environment of aeromedical evacuation. This monitoring can be particularly useful during the strategic evacuation of patients with severe injury from field hospital to tertiary centers. Care management study, level V.Keywords
This publication has 12 references indexed in Scilit:
- Comparison of monitoring performance of Bioreactance vs. pulse contour during lung recruitment maneuversCritical Care, 2009
- Evaluation of a noninvasive continuous cardiac output monitoring system based on thoracic bioreactanceAmerican Journal of Physiology-Heart and Circulatory Physiology, 2007
- Nicht-invasive Bestimmung des Herzzeitvolumens durch einen transkutanen CW Doppler in der LuftrettungAINS - Anästhesiologie · Intensivmedizin · Notfallmedizin · Schmerztherapie, 2005
- The provision of sophisticated critical care beyond the hospital: Lessons from physiology and military experiences that apply to civil disaster medical responseCritical Care Medicine, 2005
- Autonomic control of the cardiovascular system during acclimatization to high altitude: effects of sildenafilJournal of Applied Physiology, 2004
- Cardiovascular autonomic modulation and activity of carotid baroreceptors at altitudeClinical Science, 1998
- Effects of high altitude acclimatization on heart rate variability in resting humansEuropean Journal of Applied Physiology, 1996
- Heart rate and respiratory rhythm dynamics on ascent to high altitude.Heart, 1995
- Reduction of stroke volume during exercise in man following ascent to 3,100 m altitude.Journal of Applied Physiology, 1967
- Blood volume and haemoglobin concentration at altitudes above 18,000 ft. (5500 m)The Journal of Physiology, 1964