Abstract
Plate breakage is one form of construct failure after a clavicle fracture treated with an open reduction and plate fixation. A recent study evaluated construct failure after an open reduction and plate fixation and reported a construct failure rate of 6.9% of which 1.9% were related to broken plates. Plate breakage is rare, thus, there are insufficient data regarding risk factors, pathogenesis, or how to avoid it. This case report presents an unusual case of a 35-year-old Caucasian man, 7 weeks after open reduction and internal plate fixation of a fracture in the middle third of his clavicle, who developed breakage of the implant. Surgery was advised, the implant was retrieved, the fracture was reduced, and a new bridging locking plate was implanted. In the current case it seems that the use of a bridging plate, the fundamental anatomical structure of the clavicle and the forces that are applied on it, the lack of discipline in complying with the postoperative functional restrictions, and an unclear “patient expectation” process were the main reasons for the failure. These aspects should be carefully considered and addressed in clavicle fractures.Keywords
This publication has 15 references indexed in Scilit:
- Frequency and complications after operative fixation of clavicular fracturesJournal of Shoulder and Elbow Surgery, 2016
- Reconstruction plates for midshaft clavicular fractures: A retrospective cohort studyOrthopaedics & Traumatology: Surgery & Research, 2016
- Does plate type influence the clinical outcomes and implant removal in midclavicular fractures fixed with 2.7-mm anteroinferior plates? A retrospective cohort studyJournal of Orthopaedic Surgery and Research, 2014
- Patient expectations and patient-reported outcomes in surgery: A systematic reviewSurgery, 2014
- Open Reduction and Plate Fixation Versus Nonoperative Treatment for Displaced Midshaft Clavicular FracturesJournal of Bone and Joint Surgery, 2013
- Risk factors for postoperative complications of displaced clavicular midshaft fractures2012
- Treatment of clavicle fractures: current concepts reviewJournal of Shoulder and Elbow Surgery, 2012
- Treatment of Clavicle Fractures2012
- Systematic review of the complications of plate fixation of clavicle fracturesArchives of orthopaedic and trauma surgery, 2012
- Treatment of Acute Midshaft Clavicle Fractures: Systematic Review of 2144 FracturesJournal of Orthopaedic Trauma, 2005