Is it safe to inject contrast through the side arm of an introducer sheath? An in vitro study
- 1 June 2020
- journal article
- research article
- Published by Springer Nature in Emergency Radiology
- Vol. 27 (3) , 303-306
Abstract
Purpose Occasionally, a patient requiring computed tomography (CT) with poor venous access or in an unstable condition may have a sheath as their only form of IV access. Many institutions prohibit contrast power injection through the side ports, for concern for rupture. We hypothesize that under conditions encountered in CT scanning, the side arm of the sheath introducer is safe for power injection of contrast material. Methods In this in vitro study, we injected contrast at different flow rates varying from 2 to 8 ml/s, through the side port of the sheath. Time-pressure graphs were obtained for each injection from the injector display. The assembly was observed for any signs of leakage or material failure. Results There was not a single event of leak, rupture, or displacement. Maximum pressures were within the range of conventional contrast administrations. Conclusion This study suggests that iodinated contrast can safely be injected through the side port of a venous sheath.Keywords
This publication has 8 references indexed in Scilit:
- MRI and CT contrast media extravasationMedicine, 2018
- A clinical pathway for the management of difficult venous accessBMC Nursing, 2017
- Systematic review of the safety and efficacy of contrast injection via venous catheters for contrast-enhanced computed tomographyEuropean Journal of Radiology Open, 2017
- Intravenous Contrast Medium Administration and Scan Timing at CT: Considerations and ApproachesRadiology, 2010
- Power Injection of Contrast Media via Peripherally Inserted Central Catheters for CTJournal of Vascular and Interventional Radiology, 2004
- Power Injection of Contrast Media Using Central Venous CathetersAmerican Journal of Roentgenology, 2001
- Increasing the Functional Gauge of the Side Port of Large Catheter Sheath IntroducersAnesthesiology, 1982