Abstract
A number of outbreaks and pseudo-outbreaks have been associated with contaminated rinse water in endoscope washer-disinfectors. Health Technical Memorandum 2030 specifies that final rinse water should be ‘bacteria-free’. In this study, results of rinse-water testing from 20 endoscopy units were reviewed over a 4-month period. Over 60% of samples were of an unsatisfactory quality (i.e. not bacteria-free) and none of the endoscopy units consistently achieved sterile water throughout the study period. Poor microbiology results caused anxiety to endoscopy staff and infection control teams who had to decide whether or not to take washer-disinfectors out of use, possibly resulting in delays to medical procedures. There was no common policy on how to react to poor results, with staff at each unit developing their own action levels. Here, it is suggested that future guidelines would be of more practical use if they specified a series of action levels of increasing severity based on the bacterial count in a water sample.Keywords
This publication has 7 references indexed in Scilit:
- Rinse water for heat labile endoscopy equipmentJournal of Hospital Infection, 2002
- Application of Environmental Sampling to Flexible Endoscope Reprocessing: The Importance of Monitoring the Rinse WaterInfection Control & Hospital Epidemiology, 2002
- An outbreak of multidrug-resistant Pseudomonas aeruginosa infection associated with contamination of bronchoscopes and an endoscope washer-disinfectorJournal of Hospital Infection, 2000
- Mycobacterium chelonae isolated from rinse water within an endoscope washer–disinfectorJournal of Hospital Infection, 2000
- Pseudoepidemic of Legionella pneumophila serogroup 6 associated with contaminated bronchoscopesJournal of Hospital Infection, 1997
- Water as a Reservoir of Nosocomial PathogensInfection Control & Hospital Epidemiology, 1997
- Endoscope decontamination: automated vs. manualJournal of Hospital Infection, 1995