A biosensor based on a film bulk acoustic resonator and biotin–avidin system for the detection of the epithelial tumor marker mucin 1
- 30 July 2015
- journal article
- research article
- Published by Royal Society of Chemistry (RSC) in RSC Advances
- Vol. 5 (81) , 66355-66359
Abstract
AlN thin film bulk acoustic resonators (FBARs) with a resonant frequency of ∼575 MHz have been fabricated to function as an epithelial tumor marker mucin 1 (MUC1) biosensor. Streptavidin was assembled on the sensitive area of FBAR. After the recognition between aptamers–AuNP conjugates and MUC1, biotin, along with the conjugates, was captured by streptavidin onto the surface of FBARs. Therefore, the target MUC1 could be sensitively detected. This biosensor exhibited a good linear relationship between the frequency shifts and the concentrations of MUC1 ranging from 30 to 500 nM, which indicated the sensitivity is about 818.6 Hz nM−1. The frequency shift remained relatively stable when the concentration of MUC1 was greater than 500 nM since the binding between MUC1 and aptamers–AuNP conjugates reached saturation. The selectivity experiment demonstrated that this biosensor can precisely detect MUC1 with good specificity. The positive results suggest that FBAR is an attractive alternative to a new approach for the detection of MUC1.Keywords
This publication has 24 references indexed in Scilit:
- Ag Nanocluster/DNA Hybrids: Functional Modules for the Detection of Nitroaromatic and RDX ExplosivesNano Letters, 2014
- Cell-Based Biosensors and Their Application in BiomedicineChemical Reviews, 2014
- Piezotronic Effect on ZnO Nanowire Film Based Temperature SensorACS Applied Materials & Interfaces, 2014
- High sensitivity acoustic wave AlN/Si mass detectors arrays for artificial olfactory and biosensing applications: A reviewSensors and Actuators B: Chemical, 2013
- AlN-based BAW resonators with CNT electrodes for gravimetric biosensingSensors and Actuators B: Chemical, 2011
- A Nanoparticle Autocatalytic Sensor for Ag+ and Cu2+ Ions in Aqueous Solution with High Sensitivity and Selectivity and Its Application in Test PaperAnalytical Chemistry, 2011
- AlN-based sputter-deposited shear mode thin film bulk acoustic resonator (FBAR) for biosensor applications — A reviewSurface and Coatings Technology, 2010
- Detection of carcinoembryonic antigen using AlN FBARThin Solid Films, 2010
- A contour-mode film bulk acoustic resonator of high quality factor in a liquid environment for biosensing applicationsApplied Physics Letters, 2010
- Shear mode FBARs as highly sensitive liquid biosensorsSensors and Actuators A: Physical, 2006