Image quality and radiologists’ subjective acceptance using model-based iterative and deep learning reconstructions as adjuncts to ultrahigh-resolution CT in low-dose contrast-enhanced abdominopelvic CT: phantom and clinical pilot studies
Open Access
- 16 December 2021
- journal article
- research article
- Published by Springer Nature in Abdominal Radiology
- Vol. 47 (2) , 891-902
Abstract
Purpose: In contrast-enhanced abdominopelvic CT (CE-APCT) for oncologic follow-up, ultrahigh-resolution CT (UHRCT) may improve depiction of fine lesions and low-dose scans are desirable for minimizing the potential adverse effects by ionizing radiation. We compared image quality and radiologists’ acceptance of model-based iterative (MBIR) and deep learning (DLR) reconstructions of low-dose CE-APCT by UHRCT. Methods: Using our high-resolution (matrix size: 1024) and low-dose (tube voltage 100 kV; noise index: 20–40 HU) protocol, we scanned phantoms to compare the modulation transfer function and noise power spectrum between MBIR and DLR and assessed findings in 36 consecutive patients who underwent CE-APCT (noise index: 35 HU; mean CTDIvol: 4.2 ± 1.6 mGy) by UHRCT. We used paired t-test to compare objective noise and contrast-to-noise ratio (CNR) and Wilcoxon signed-rank test to compare radiologists’ subjective acceptance regarding noise, image texture and appearance, and diagnostic confidence between MBIR and DLR using our routine protocol (matrix size: 512; tube voltage: 120 kV; noise index: 15 HU) for reference. Results: Phantom studies demonstrated higher spatial resolution and lower low-frequency noise by DLR than MBIR at equal doses. Clinical studies indicated significantly worse objective noise, CNR, and subjective noise by DLR than MBIR, but other subjective characteristics were better (P < 0.001 for all). Compared with the routine protocol, subjective noise was similar or better by DLR, and other subjective characteristics were similar or worse by MBIR. Conclusion: Image quality, except regarding noise characteristics, and acceptance by radiologists were better by DLR than MBIR in low-dose CE-APCT by UHRCT. GraphicalKeywords
Funding Information
- The Japan Society for the Promotion of Science, Japan ((C) 18k07643)
This publication has 36 references indexed in Scilit:
- Standard dose versus low-dose abdominal and pelvic CT: Comparison between filtered back projection versus adaptive iterative dose reduction 3DDiagnostic and Interventional Imaging, 2014
- Incidence of contrast-induced nephropathy in hospitalised patients with cancerEuropean Radiology, 2013
- Physician Preference Between Low-Dose Computed Tomography With a Sinogram-Affirmed Iterative Reconstruction Algorithm and Routine-Dose Computed Tomography With Filtered Back Projection in Abdominopelvic ImagingJournal of Computer Assisted Tomography, 2013
- Dose reduction in oncological staging multidetector CT: effect of iterative reconstructionBritish Journal of Radiology, 2013
- Contrast induced nephropathy: updated ESUR Contrast Media Safety Committee guidelinesEuropean Radiology, 2011
- Cancer and renal insufficiency results of the BIRMA studyBritish Journal of Cancer, 2010
- Estimating Effective Dose for CT Using Dose–Length Product Compared With Using Organ Doses: Consequences of Adopting International Commission on Radiological Protection Publication 103 or Dual-Energy ScanningAmerican Journal of Roentgenology, 2010
- Projected Cancer Risks From Computed Tomographic Scans Performed in the United States in 2007Archives of internal medicine (1960), 2009
- Iterative Reconstruction Technique for Reducing Body Radiation Dose at CT: Feasibility StudyAmerican Journal of Roentgenology, 2009
- Abdominal Helical CT: Evaluation of Optimal Doses of Intravenous Contrast Material—A Prospective Randomized StudyRadiology, 2000