Analytical Modeling of Small, Solenoidal, and Implantable Coils With Ferrite Tube Core
- 24 January 2019
- journal article
- research article
- Published by Institute of Electrical and Electronics Engineers (IEEE) in IEEE Microwave and Wireless Components Letters
- Vol. 29 (3) , 237-239
Abstract
Implantable biomedical devices are required to be compact in size for reducing the harmful impact on human tissue, which poses a significant challenge for energizing the devices wirelessly. Inserting a ferrite tube into a solenoidal power receiving coil can improve the coupling between the coils for wireless power transfer yet keep the device compact. Based on the analytical effective permeability model of the ferrite tube, an analytical model of the solenoidal coil with the ferrite tube core is established in this letter. Compared to the existing models, the accuracy is significantly improved by using this new model, which is validated by both simulation and measurement in air and tissue environments.Keywords
Funding Information
- National Natural Science Foundation of China (61771175)
- Natural Science Foundation of Zhejiang Province (LY17F010018)
This publication has 13 references indexed in Scilit:
- The design of high efficiency energy receiving coil for micro-ball Endoscopy2016
- Ultra-Thin Wireless Power Module with Integration of Wireless Inductive Link and Supercapacitors2016
- Characterization of Two-Turns External Loop Antenna With Magnetic Core for Efficient Wireless Powering of Cortical ImplantsIEEE Antennas and Wireless Propagation Letters, 2015
- A capsule robot powered by wireless power transmission: Design of its receiving coilSensors and Actuators A: Physical, 2015
- Enhancement of wireless power transfer efficiency using flat conductor with ferrite wallMicrowave and Optical Technology Letters, 2015
- Wireless Power Transfer for Telemetric Devices With Variable Orientation, for Small Rodent Behavior MonitoringIEEE Sensors Journal, 2014
- An Analytical Model for Inductively Coupled Implantable Biomedical Devices With Ferrite RodsIEEE Transactions on Biomedical Circuits and Systems, 2008
- An inductive power link for a wireless endoscopeBiosensors and Bioelectronics, 2007
- BION ™ system for distributed neural prosthetic interfacesMedical Engineering & Physics, 2001
- The dielectric properties of biological tissues: II. Measurements in the frequency range 10 Hz to 20 GHzPhysics in Medicine & Biology, 1996