Trends in maintenance status and usability of public automated external defibrillators during a 5-year on-site inspection
Authors:
Open Access
- 24 June 2022
- journal article
- research article
- Published by Springer Nature in Scientific Reports
- Vol. 12 (1) , 1-8
Abstract
This study aimed to assess the trend of the maintenance status and usability of public automated external defibrillators (AEDs). Public AEDs installed in Seoul from 2013 to 2017 were included. An inspector checked the maintenance status and usability of the AEDs annually using a checklist. During the study period, 23,619 AEDs were inspected. Access to the AEDs was improved, including the absence of obstacles near the AEDs (from 90.2% in 2013 to 99.1% in 2017, p < 0.0001) and increased AED signs (from 34.3% in 2013 to 91.3% in 2017, p < 0.0001). The rate of AEDs in normal operation (from 94.0% in 2013 to 97.6% in 2017, p < 0.0001), good battery status (from 95.6% in 2013 to 96.8% in 2017, p = 0.0016), and electrode availability increased (from 97.1% in 2013 to 99.0% in 2017, p < 0.0001); the rate of electrode validity decreased (from 90.0% in 2013 to 87.2% in 2017, p < 0.0001). The overall rate of the non-ready-to-use AEDs and AEDs with less than 24-h usability accounted for 15.4% and 44.1% of the total number of AEDs, respectively. Although most AEDs had a relatively good maintenance status, a significant proportion of public AEDs were not available for 24-h use. Invalid electrodes and less than 24-h accessibility were the main reasons that limited the 24-h usability of public AEDs.Keywords
All Related Versions
This publication has 20 references indexed in Scilit:
- Public Access Defibrillation: Great benefit and potential but infrequently usedResuscitation, 2015
- Part 3: Adult basic life support and automated external defibrillationResuscitation, 2015
- Automated External Defibrillators Inaccessible to More Than Half of Nearby Cardiac Arrests in Public Locations During Evening, Nighttime, and WeekendsCirculation, 2013
- Analysis of Automated External Defibrillator Device Failures Reported to the Food and Drug AdministrationAnnals of Emergency Medicine, 2012
- Recalls and Safety Alerts Affecting Automated External DefibrillatorsJAMA, 2006
- Community Lay Rescuer Automated External Defibrillation ProgramsCirculation, 2006
- Self-reported checklist logs overestimate compliance with defibrillator inspectionsThe Quality Assurance Journal, 2005
- Public-Access Defibrillation and Survival after Out-of-Hospital Cardiac ArrestNew England Journal of Medicine, 2004
- ESC-ERC recommendations for the use of automated external defibrillators (AEDs) in EuropeEuropean Heart Journal, 2004
- Tripling Survival From Sudden Cardiac Arrest Via Early Defibrillation Without Traditional Education in Cardiopulmonary ResuscitationCirculation, 2002