A portable OCT scanner usable for dental applications is designed to be compact (B4 notebook size), lightweight, and capable of DC battery operation, enabling chairside use and mobility in home medical care. TD-OCT faces the challenge of a slower measurement speed compared to SD/SS-OCT. In this study, by utilizing multiple rotating reflectors and combining them with a 3 × 3 fiber coupler, a measurement range of over 10 mm and a measurement speed of 1000 scans per second were achieved. Additionally, the dental intraoral probe was designed to ensure high accessibility, allowing free positioning within the oral cavity, with an ingeniously compact optical system. A cylindrical probe with a diameter of 6 mmϕ and a length of 10 mm was developed to enable tooth measurement. This study demonstrated the capability to evaluate attenuation coefficients derived from material properties, starting with measurements of caries and gums.Keywords
This publication has 35 references indexed in Scilit:
- Study on an all-fiber optical coherence tomography (OCT) probeOptical Fiber Technology, 2023
- Micro crack analysis of optical fiber by specialized TD-OCTOptics & Laser Technology, 2019
- Optical coherence tomography assessment of gingival epithelium inflammatory status in periodontal — Systemic affected patientsAnnals of Anatomy - Anatomischer Anzeiger, 2018
- Application of Optical Coherence Tomography (OCT) for Diagnosis of Caries, Cracks, and Defects of RestorationsCurrent Oral Health Reports, 2015
- Clinical assessment of non carious cervical lesion using swept‐source optical coherence tomographyJournal of Biophotonics, 2014
- Noninvasive cross‐sectional imaging of proximal caries using swept‐source optical coherence tomography (SS‐OCT) in vivoJournal of Biophotonics, 2013
- Effects of structural orientation of enamel and dentine on light attenuation and local refractive index: An optical coherence tomography studyJournal of Dentistry, 2012
- Validation of swept-source optical coherence tomography (SS-OCT) for the diagnosis of occlusal cariesJournal of Dentistry, 2010
- In vivo diagnosis of oral dysplasia and malignancy using optical coherence tomography: Preliminary studies in 50 patientsLasers in Surgery and Medicine, 2009
- Performance of a new laser fluorescence device for the detection of occlusal caries in vitroJournal of Dentistry, 2006