The comparison of thermal tissue injuries caused by ultrasonic scalpel and electrocautery use in rabbit tongue tissue
Open Access
- 20 August 2012
- journal article
- research article
- Published by Association of Basic Medical Sciences of FBIH in Biomolecules and Biomedicine
- Vol. 12 (3) , 151-157
Abstract
The aim of this study compares to the increase in tissue temperature and the thermal histological effects of ultrasonic scalpel, bipolar and unipolar electrosurgery incisions in the tongue tissue of rabbits. This study evaluates the histopathological changes related to thermal change and the maximum temperature values in the peripheral tissue brought about by the incisions carried out by the three methods in a comparative way. To assess thermal tissue damage induced by the three instruments, maximum tissue temperatures were measured during the surgical procedure and tongue tissue samples were examined histopathologically following the surgery. The mean maximum temperature values of the groups were 93.93±2.76 C° for the unipolar electrocautery group, whereas 85.07±5.95 C° for the bipolar electrocautery group, and 108.23±7.64 C° for the ultrasonic scalpel group. There was a statistically significant relationship between the increase in maximum temperature values and the separation among tissue layers, edema, congestion, necrosis, hemorrhage, destruction in blood vessel walls and fibrin accumulation, and between the existence of fibrin thrombus and tissue damage depth (p<0.05). It was concluded that the bipolar electrocautery use gives way to less temperature increase in the tissues and less thermal tissue damage in comparison to the other methods.Keywords
This publication has 15 references indexed in Scilit:
- Electrocautery: Principles and PracticeJournal of Endourology, 2003
- Determination of temperature elevation in tissue during the application of the harmonic scalpelUltrasound in Medicine & Biology, 2003
- Bipolar versus monopolar electrosurgery: clinical applications.2001
- Comparison of monopolar and bipolar electrosurgical modes for restorative dentistry: A review of the literatureThe Journal of Prosthetic Dentistry, 2001
- Electrosurgery in Otolaryngology–Head and Neck Surgery: Principles, Advances, and ComplicationsThe Laryngoscope, 2001
- Reflection-optical multispectral imaging method for objective determination of burn depthBurns, 1999
- A comparison of mucosal incisions made by scalpel, CO laser, electrocautery, and constant-voltage electrocauteryOtolaryngology–Head and Neck Surgery, 1997
- Thermal Injury Patterns and Tensile Strength of Canine Oral Mucosa After Carbon Dioxide Laser IncisionsThe Laryngoscope, 1996
- The plasma scalpel: A new thermal knifeLasers in Surgery and Medicine, 1982
- Histologic evaluation of electrosurgery with varying frequency and waveformThe Journal of Prosthetic Dentistry, 1978