Performance evaluation of the digital cell imaging analyzer DI-60 integrated into the fully automated Sysmex XN hematology analyzer system
- 1 January 2015
- journal article
- research article
- Published by Walter de Gruyter GmbH in Clinical Chemistry and Laboratory Medicine (CCLM)
- Vol. 53 (2) , 281-289
Abstract
Background: The XN-Series (Sysmex, Kobe, Japan) have been equipped with the automated digital cell imaging analyzer DI-60, which provides complete automation of the sample processing with automated complete blood counts (CBC), slide making/staining, and digital scanning with cell pre-classification. The aim of this study was to evaluate the efficacy of the XN-Series as an integrated blood cell analysis system. Methods: White blood cell (WBC) morphological analysis by the DI-60 was evaluated using 232 blood samples from patients. Routine analysis of a total of 2000 blood samples has been performed to evaluate the processing ability of the XN-Series connected to the DI-60. Results: The overall analysis accuracy of pre-classification of WBC by the DI-60 was 88.4%. Good correlation was observed between final results of the DI-60 analysis and manual differentiation with high sensitivity and specificity for blasts and immature granulocytes. The sample processing time of the XN-Series, from automated CBC to cell pre-classification, was 38±1 min/single run and 165±12 min/500 CBC samples run (slide preparation rate 15.6%) with no sample hold-up at the DI-60. Conclusions: The automated morphological analysis capability of the DI-60 has potential usefulness in the integrated automated hematology analysis system of XN-Series.Keywords
This publication has 11 references indexed in Scilit:
- Validation of the Sysmex sp-1000i automated slide preparer-stainer in a clinical laboratoryRevista Brasileira de Ginecologia e Obstetrícia, 2013
- Performance evaluation of the Sysmex haematology XN modular systemJournal of Clinical Pathology, 2012
- Improved Flagging Rates on the Sysmex XE-5000 Compared With the XE-2100 Reduce the Number of Manual Film Reviews and Increase Laboratory ProductivityAmerican Journal of Clinical Pathology, 2011
- Clinical Practice Guideline for the Use of Antimicrobial Agents in Neutropenic Patients with Cancer: 2010 Update by the Infectious Diseases Society of AmericaClinical Infectious Diseases, 2011
- ORIGINAL ARTICLE: Evaluation of the CellaVision™ DM automated microscope in pediatricsInternational Journal of Laboratory Hematology, 2010
- Performance of automated slidemakers and stainers in a working laboratory environment – routine operation and quality controlInternational Journal of Laboratory Hematology, 2010
- Can automated blood film analysis replace the manual differential? An evaluation of the CellaVision DM96 automated image analysis systemInternational Journal of Laboratory Hematology, 2009
- Examination of peripheral blood films using automated microscopy; evaluation of Diffmaster Octavia and Cellavision DM96Journal of Clinical Pathology, 2006
- Performance Evaluation of the CellaVision DM96 SystemAmerican Journal of Clinical Pathology, 2005
- Automated differential leucocyte counting by the Geometric Data Hematrak system: Eighteen months experience in a private pathology laboratoryPathology, 1980