Measurements of computed tomography x-ray fields utilizing the partial volume effect

Authors:
  • RJRobert A. Jucius
    Robert A. Jucius
  • GKGeorge X. Kambic
    George X. Kambic

Abstract
Due to the X-ray field geometry, methods to determine from computed tomographic (CT) scanners require modification from standard techniques. Ionization chambers have geometries which do not allow measurement of narrow, rotating X-ray fields. By utilizing the partial volume effect, an ion chamber can be calibrated to measure narrow X-ray fields with a spatial resolution of down to 1 mm. The number of scans required to reconstruct a complete profile depends on the resolution required. It is typically around 10-20 scans. This is accomplished using a series of attenuating sleeves over the ionization chamber. The chamber and sleeves are calibrated against a standard chamber using large field geometries. Comparisons with TLD measurements verify the accuracy of the technique.