Automated Cerebrospinal Fluid Cell Counts Using the Sysmex XE-5000
Authors:
- Pedro CiarliniPCPedro Ciarlini
- Diane KidricDKDiane Kidric
- Christine DillmanCDChristine Dillman
- MaryAnn O’RiordanMOMaryAnn O’Riordan
- 1 November 2010
- journal article
- research article
- Published by Oxford University Press (OUP) in American Journal of Clinical Pathology
- Vol. 134 (5) , 734-738
Abstract
The main objectives of the study were to compare manual and automated WBC counts on clear cerebrospinal fluid (CSF) samples. Clear CSF samples from 200 adults and children were studied. Cell counts were performed manually using a hemocytometer and then analyzed on the Sysmex XE-5000. Descriptive statistics and Spearman correlation for nonparametric data were used for method comparison. Manual WBC counts ranged from 0 to 702 cells/μL, and Sysmex counts ranged from 0 to 629 cells/μL. The Spearman rank correlation coefficient for the entire range of data was 0.77 (P < .001); however, the correlation was weaker at the low end of the data spectrum. For manual WBC ranges of 0 to 5 cells/μL and 0 to 10 cells/μL, the corresponding Sysmex 0 to 95th percentile ranges were 0 to 23 cells/μL and 0 to 27 cells/μL, respectively. The results suggest that larger studies are necessary to determine new reference ranges for automated CSF WBC counts.Keywords
This publication has 13 references indexed in Scilit:
- Establishing Reference Intervals for Clinical Laboratory Test ResultsAmerican Journal of Clinical Pathology, 2010
- Estimating Reference IntervalsAmerican Journal of Clinical Pathology, 2010
- Evaluation of the XE-5000 for the automated analysis of blood cells in cerebrospinal fluidClinical Biochemistry, 2009
- Comparative Evaluation of the Iris iQ200 Body Fluid Module With Manual Hemacytometer CountAmerican Journal of Clinical Pathology, 2009
- The Clinical Reliability of Automated Cerebrospinal Fluid Cell Counts on the Beckman-Coulter LH750 and Iris iQ200American Journal of Clinical Pathology, 2009
- Automated counting of cells in cerebrospinal fluidInternational Journal of Laboratory Hematology, 2005
- An Evaluation of the Cell-Dyn 3200 for Counting Cells in Cerebrospinal and Other Body FluidsLaboratory Hematology, 2005
- Performance Evaluation of the Application of Body Fluids on the Sysmex XE-2100 Series Automated Hematology AnalyzerLaboratory Hematology, 2005
- Evaluation of ADVIA 120 CSF assay (BayerR) vs. chamber counting of cerebrospinal fluid specimensClinical and Laboratory Haematology, 2004
- Book Review: Clinical Diagnosis and Management by Laboratory MethodsLaboratory Medicine, 2001