Accuracy of the Withings BPM Pro 2 (WPM07) Device for Self-Blood Pressure Measurements in the General Populations According to the International Organization for Standardization Universal Standard (ISO 81060-2:2018/AMD 2:2024) Protocol
Open Access
- 1 February 2026
- journal article
- research article
- Published by Taylor & Francis in Vascular Health and Risk Management
- Vol. ume 22, 1-9
Abstract
Background: The accuracy of blood pressure (BP) measurements performed by electronic devices is crucial in clinical practice. International professional societies emphasize the need for rigorous, standardized validation of electronic BP measuring devices to ensure their accuracy and reliability. Objective: This study evaluated the accuracy of the Withings BPM Pro2 (WPM07) upper‑arm oscillometric BP monitor in the adult general population using the ISO 81060‑2:2018/AMD 2:2024 Universal Standard. Methods: Following the ISO same‑arm sequential measurement protocol, 89 eligible pre-selected participants underwent paired test‑device and mercury reference auscultatory measurements. Nine BP measurements were performed on each participant, four with the tested device and five with the auscultatory method. The accuracy of BP measurements was assessed based on the difference between simultaneous measurements taken by two observers and those displayed by the tested device using ISO Criterion 1 (mean and standard deviation of test–reference differences) and Criterion 2 (within‑subject variability of averaged difference values). Results: Eighty-nine participants underwent the BP measurements protocol, data from 87 were analyzed after excluding 2 participants due to excessive BP variability. Inter‑observer agreement for reference measurements was 0.64 ± 2.14 mmHg for systolic BP (SBP) and 0.39 ± 1.98 mmHg for diastolic BP (DBP). ISO criterion 1 was met with mean test–reference differences of 1.0 ± 4.9 mmHg (SBP) and 0.6 ± 4.1 mmHg (DBP) below ISO requirements (≤ 5± 8 mmHg). ISO criterion 2 was also satisfied, with within‑subject standard deviations of 3.50 mmHg (SBP) and 3.01 mmHg (DBP), well below the ISO thresholds (≤ 6.87/6.91 mmHg). Conclusion: The BPM Pro2 (WPM07) fulfills the ISO 81060‑2:2018/AMD 2:2024 accuracy requirements and is suitable for reliable home BP self‑measurement in adults.Keywords
This publication has 16 references indexed in Scilit:
- 2023 ESH Guidelines for the management of arterial hypertension The Task Force for the management of arterial hypertension of the European Society of HypertensionJournal Of Hypertension, 2023
- 2021 European Society of Hypertension practice guidelines for office and out-of-office blood pressure measurementJournal Of Hypertension, 2021
- Lancet Commission on Hypertension group position statement on the global improvement of accuracy standards for devices that measure blood pressureJournal Of Hypertension, 2020
- Can an automatic oscillometric device replace a mercury sphygmomanometer on blood pressure measurement? a systematic review and meta-analysisBlood Pressure Monitoring, 2019
- Self‐blood pressure monitoring as a tool to increase hypertension awareness, adherence to antihypertensive therapy, and blood pressure controlThe Journal of Clinical Hypertension, 2019
- Self‐blood pressure monitoring is associated with improved awareness, adherence, and attainment of target blood pressure goals: Prospective observational study of 7751 patientsThe Journal of Clinical Hypertension, 2019
- Recommendations and Practical Guidance for performing and reporting validation studies according to the Universal Standard for the validation of blood pressure measuring devices by the Association for the Advancement of Medical Instrumentation/European Society of Hypertension/International Organization for Standardization (AAMI/ESH/ISO)Journal Of Hypertension, 2019
- A Universal Standard for the Validation of Blood Pressure Measuring DevicesHypertension, 2018
- European Society of Hypertension International Protocol revision 2010 for the validation of blood pressure measuring devices in adultsBlood Pressure Monitoring, 2010
- Short report: An outline of the revised British Hypertension Society protocol for the evaluation of blood pressure measuring devicesJournal Of Hypertension, 1993