Impact of belly board immobilization devices and body mass factor on setup displacement using daily cone‐beam CT in rectal cancer radiotherapy
Authors:
Open Access
- 29 November 2024
- journal article
- research article
- Published by Wiley in Journal of Applied Clinical Medical Physics
- Vol. 26 (3) , e14573
Abstract
Objective: The objective of this study is to evaluate the impact of different belly board and daily changes in patient's body‐mass factor (BMF) on setup displacement in radiotherapy for rectal cancer.Methods: Twenty‐five patients were immobilized using the thermoplastic mask with belly board (TM‐BB), and 30 used the vacuum bag cushion with belly board (VBC‐BB), performing daily cone‐beam computed tomography (CBCT) scans 625 times and 750 times, respectively. Daily pretreatment CBCT scans were registered to the planned CT images for BMF change determination and setup displacement measurement. Independent t‐tests compared setup displacement between the two groups in left‐right (LR), superior‐inferior (SI), and anterior‐posterior (AP) directions, as well as the BMF changes. The impact of daily BMF changes on setup displacement was evaluated using multivariate logistic regression and 10‐fold cross‐validation.Results: The setup displacement for TM‐BB in the LR, SI, and AP directions were 0.31 ± 0.25, 0.58 ± 0.40, and 0.19 ± 0.18 cm, respectively, while VBC‐BB showed 0.19 ± 0.15, 0.26 ± 0.22, and 0.36 ± 0.29 cm in the corresponding directions, respectively. Margins of planning target volume (PTV) for TM‐BB were 8, 10, and 6 mm in LR, SI, and AP directions, while VBC‐BB showed margins of 5,7, and 8 mm, respectively. The daily BMF changes for both groups were ranked in descending order as follows: sacral rotation angle (RS), hip lateral diameter (HLD), and hip anterior‐posterior diameter (HAPD). HAPD was the main factor affecting setup displacement in both the AP and SI directions in TM‐BB, while RS was the primary factor for setup displacement in the AP direction in VBC‐BB.Conclusion: Compared with TM‐BB, VBC‐BB had a larger AP displacement but smaller in LR and SI displacement. Daily changes in BMF have distinct effects on setup displacement in different immobilization devices. Image‐guided radiation therapy (IGRT) is highly recommended and BMF changes should be given consideration during radiotherapy.Keywords
This publication has 38 references indexed in Scilit:
- Effect of Body Mass Factors on Setup Displacement in Gynecologic Tumors and Subsequent Effect on PTV MarginsAdvances in Radiation Oncology, 2022
- Factors Affecting Isocenter Displacement and Planning Target Volume Margin for Patients With Rectal Cancer Receiving Radiation TherapyAdvances in Radiation Oncology, 2022
- Adoption of Total Neoadjuvant Therapy for Locally Advanced Rectal CancerJAMA Oncology, 2018
- Setup errors in radiation therapy for thoracic tumor patients of different body mass indexJournal of Applied Clinical Medical Physics, 2018
- Planning Target Volume Margin Evaluation and Critical Structure Sparing for Rectal Cancer Patients Treated Prone on a BellyboardClinical Oncology, 2013
- Image-guided Radiotherapy for Rectal Cancer — A Systematic ReviewClinical Oncology, 2012
- Quantification of Organ Motion During Chemoradiotherapy of Rectal Cancer Using Cone-Beam Computed TomographyInternational Journal of Radiation Oncology - Biology - Physics (IJROBP), 2011
- Evaluation of a Thermoplastic Immobilization System for Breast and Chest Wall Radiation TherapyMedical Dosimetry, 2011
- Rotation of the Sacrum During Bellyboard Pelvic RadiotherapyMedical Dosimetry, 2010
- Elective Clinical Target Volumes for Conformal Therapy in Anorectal Cancer: A Radiation Therapy Oncology Group Consensus Panel Contouring AtlasInternational Journal of Radiation Oncology - Biology - Physics (IJROBP), 2009