Abstract
The audiogram predicts R 2  = .31) and STM sensitivity ( R 2  = .28). Combined with the HFA, STM sensitivity significantly improved the SRT prediction (Δ R 2  = .13; total R 2  = .44). The remaining unaccounted variance might be attributable to variability in cognitive function and other dimensions of suprathreshold distortion. STM sensitivity was most critical in predicting SRTs for listeners < 65 years old or with HFA <53 dB HL. Results are discussed in the context of previous work suggesting that STM sensitivity for low rates and low-frequency carriers is impaired by a reduced ability to use temporal fine-structure information to detect dynamic spectra. STM detection is a fast test of suprathreshold auditory function for frequencies <2 kHz that complements the HFA to predict variability in hearing-aid outcomes for speech perception in noise.

This publication has 67 references indexed in Scilit: