Measurement of detection efficiency and response functions for an Si(Li) x‐ray spectrometer in the range 0.1–5 keV
Authors:
- 14 March 2001
- journal article
- research article
- Published by Wiley in X-Ray Spectrometry
- Vol. 30 (2) , 69-76
Abstract
Efficiency and response functions were determined for a modern commercial energy‐dispersive x‐ray spectrometer in the photon energy range 0.1–5 keV using dispersed and undispersed synchrotron radiation. The spectrometer was equipped with an Si(Li) detector crystal, a thin‐film window and a digital pulse processor. Monochromatized synchrotron radiation from the PTB VUV radiometry beamline at the electron storage ring BESSY was used for the absolute determination of the detection efficiency with a typical relative uncertainty of 1–2% in the photon energy range 0.1–1.5 keV by direct comparison with calibrated photodiodes. At higher photon energies, the efficiency was found by comparison of the measured and calculated undispersed synchrotron radiation spectrum. The absolute intensity of the synchrotron radiation was known with a relative uncertainty of less than 1%. In the overlapping region, the results from the two independent experiments are in full agreement. The energy dependence of the measured efficiency can be explained only with a detector model assuming that there is no dead layer. A simple model for the effect of incomplete charge collection (ICC) was applied to describe the measured response functions. Consequences of the ICC such as broadening and shift of low‐energy peaks and redistribution of counts around 1.84 keV are explained with the model, in accordance with the experimental results. Copyright © 2001 John Wiley & Sons, Ltd.Keywords
This publication has 22 references indexed in Scilit:
- High-accuracy calibration of the HXDS HPGe detector at the PTB radiometry laboratory at BESSY1997
- Characterising a Si(Li) detector element for the SIXA X-ray spectrometerNuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, 1997
- The Effect of Detector Dead Layers on Light Element Detection1995
- Modeling the Energy Dispersive X-Ray Detector1995
- Characterization of a windowless Si(Li) detector in the photon energy range 0.1–5 keVNuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, 1994
- Response functions of a Si(Li) detector for photon energies from 1 to 10 keVNuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, 1993
- Electron storage ring BESSY as a source of calculable spectral photon flux in the x-ray regionReview of Scientific Instruments, 1992
- Monte Carlo simulation of the response of Si(Li) X-ray detectors to proton induced K X-rays of light elements (12 ≦ Z ≦ 32) applied to efficiency determinationNuclear Instruments and Methods in Physics Research Section B: Beam Interactions with Materials and Atoms, 1987
- Analytic description of Si(Li) spectral lineshapes due to monoenergetic photonsX-Ray Spectrometry, 1987
- Accurate calibration of a Si(Li) detector for PIXE analysisNuclear Instruments and Methods in Physics Research Section B: Beam Interactions with Materials and Atoms, 1984