A single-center experience on long-term clinical performance of a rapid SARS-CoV-2 Antigen Detection Test, STANDARD Q COVID-19 Ag Test
Authors:
Open Access
- 27 November 2023
- journal article
- research article
- Published by Springer Nature in Scientific Reports
- Vol. 13 (1) , 1-7
Abstract
The COVID-19 pandemic in Korea has dynamically changed with the occurrence of more easily transmissible variants. A rapid and reliable diagnostic tool for detection of SARS-CoV-2 is needed. While RT-PCR is currently the gold standard for detecting SARS-CoV-2, the procedure is time-consuming and requires expert technicians. The rapid antigen detection test (RADT) was approved as a confirmatory test on 14 March 2022 due to rapid dissemination of the Omicron variant. The benefits of the RADT are speed, simplicity, and point-of-care feasibility. The aim of our study was to evaluate the clinical performance of RADT compared to RT-PCR in a single center over 15 months, fully covering the SARS-CoV-2 ‘Variants of Concern (VOC).’ A total of 14,194 cases was simultaneously tested by RT-PCR and RADT from January 2021 to March 2022 in Gangnam Severance Hospital and were retrospectively reviewed. PowerChek SARS-CoV-2, Influenza A&B Multiplex Real-time PCR Kit, and STANDARD Q COVID-19 Ag Test were used. Positive rates, sensitivities, specificities, positive predictive values (PPV), and negative predictive values (NPV) were estimated for five periods (3 months/period). Receiver operator characteristic curve (ROC) analysis was performed, and Spearman’s rank test assessed the correlation between RT-PCR Ct values and semi-quantitative RADT results. The overall positive rate of RT-PCR was 4.64%. The overall sensitivity and specificity were 0.577 [95% confidence interval (CI) 0.539–0.614] and 0.991 [95% CI 0.989–0.993], respectively. ROC analysis resulted in an area under the curve of 0.786 (P < 0.0001, Yuden’s index = 0.568). The PCR positive rates were estimated as 0.11%, 0.71%, 4.51%, 2.02%, and 13.72%, and PPV was estimated as 0.045, 0.421, 0.951, 0.720, and 0.798 in Periods 1, 2, 3, 4, and 5, respectively. A significant and moderate negative correlation between PCR Ct values and semi-quantitative RADT results was observed (Spearman’s ρ = − 0.646, P < 0.0001). The RADT exhibited good performance in specimens with low Ct values (Ct ≤ 25.00) by RT-PCR. The PPV was significantly higher in Periods 3 and 5, which corresponds to rapid dissemination of the Delta and Omicron variants. The high PPV implies that individuals with a positive RADT result are very likely infected with SARS-CoV-2 and would require prompt quarantine rather than additional RT-PCR testing. The sensitivity of 0.577 indicates that RADT should not replace RT-PCR. Nonetheless, given the high PPV and the ability to track infected persons through rapid results, our findings suggest that RADT could play a significant role in control strategies for further SARS-CoV-2 variants.Keywords
This publication has 31 references indexed in Scilit:
- Clinical Evaluation of BD Veritor SARS-CoV-2 Point-of-Care Test Performance Compared to PCR-Based Testing and versus the Sofia 2 SARS Antigen Point-of-Care TestJournal of Clinical Microbiology, 2020
- Molecular Architecture of the SARS-CoV-2 VirusCell, 2020
- Current Status of Laboratory Diagnosis for COVID-19: A Narrative ReviewInfection and Drug Resistance, 2020
- US CDC Real-Time Reverse Transcription PCR Panel for Detection of Severe Acute Respiratory Syndrome Coronavirus 2Emerging Infectious Diseases, 2020
- Severe Acute Respiratory Syndrome Coronavirus 2 from Patient with Coronavirus Disease, United StatesEmerging Infectious Diseases, 2020
- Viral RNA load as determined by cell culture as a management tool for discharge of SARS-CoV-2 patients from infectious disease wardsEuropean Journal of Clinical Microbiology & Infectious Diseases, 2020
- Recent advances and perspectives of nucleic acid detection for coronavirusJournal of Pharmaceutical Analysis, 2020
- Detection of 2019 novel coronavirus (2019-nCoV) by real-time RT-PCREurosurveillance, 2020
- Initial viral load of a COVID-19-infected case indicated by its cycle threshold value of polymerase chain reaction could be used as a predictor of its transmissibility - An experience from Gujarat, IndiaIndian Journal of Community Medicine, 2020
- Clinical tests: sensitivity and specificityContinuing Education in Anaesthesia Critical Care & Pain, 2008