Radiotherapy-induced pacemaker and implantable cardioverter defibrillator malfunction
Authors:
- Daniel W NgDNDaniel W Ng
- Komandoor SrivathsanKSKomandoor Srivathsan
- Gregory T AltemoseGAGregory T Altemose
- Michele Y HalyardMHMichele Y Halyard
- Luis R ScottLSLuis R Scott
- 1 May 2009
- journal article
- review article
- Published by Taylor & Francis in Expert Review of Medical Devices
- Vol. 6 (3) , 243-249
Abstract
It is well known that ionizing radiation can interfere with circuits in permanent pacemakers and implantable cardioverter defibrillators. Contemporary implantable cardiac devices use complementary metal-oxide silicon in combination with other very sensitive transistors. These sensitive components are especially susceptible to electromagnetic and ionizing radiation, which can potentially cause permanent damage. Electromagnetic interference is, in general, a transient phenomenon. Radiologic imaging tests have been implicated in rare cases of implantable device dysfunction and these events have been mostly transient. The American Association of Physicists in Medicine last published recommendations regarding irradiation of pacemakers in 1994. This publication is outdated and may not be pertinent for the current technology used both in the field of artificial cardiac pacing and defibrillation and in the field of radiation oncology. Updated guidelines are definitely needed.Keywords
This publication has 29 references indexed in Scilit:
- Effects of Scatter Radiation on ICD and CRT FunctionPacing and Clinical Electrophysiology, 2008
- Cancer Statistics, 2008CA: A Cancer Journal for Clinicians, 2008
- Runaway Implantable Defibrillator—A Rare Complication of Radiation TherapyPacing and Clinical Electrophysiology, 2007
- Does High-Power Computed Tomography Scanning Equipment Affect the Operation of Pacemakers?Circulation Journal, 2006
- Radiotherapy to patients with artificial cardiac pacemakersCancer Treatment Reviews, 2005
- The malfunction of permanent pacemaker induced during CT scanningJapanese Journal of Electrocardiology, 2005
- Radiation‐Induced Effects in Multiprogrammable Pacemakers and Implantable DefibrillatorsPacing and Clinical Electrophysiology, 1991
- Pacemaker Failure Due to Radiation TherapyPacing and Clinical Electrophysiology, 1982
- Damaging Effect of Therapeutic Radiation on Programmable PacemakersPacing and Clinical Electrophysiology, 1982
- Inhibition of discharge of an external demand pacemaker by an electric razorThe American Journal of Cardiology, 1971