Delayed extraction time-of-flight MALDI mass spectrometry of proteins above 25000 Da

Authors:

Abstract
Matrix‐assisted laser desorption/ionization (MALDI) with delayed extraction (DE) has been optimized for mass analysis of high‐mass proteins and glycoproteins with masses above 25000 Da. Under optimized experimental conditions, i.e. using a weak extraction field strength and a long delay time, a steep drop in mass resolution above 30000 Da is no longer observed and an improvement of more than a factor of 10 is obtained compared with the non‐DE case, at least up to 66 kDa in a 1.2 m time‐of‐flight mass analyzer. On this level of resolution the factors limiting further improvements become apparent, i.e. adduct ion formation between matrix and analyte, but also cationization and further non‐matrix‐related adducts, as well as prompt fragmentation. Moreover, heterogeneity of the sample is often the reason for the detection of broad signals for larger proteins. Within these limitations, DHBs (a 9:1 mixture of 2,5‐dihydroxybenzoic acid and 2‐hydroxy‐5‐methoxybenzoic acid) gave by far the best results. © 1997 John Wiley & Sons, Ltd.