Apparent Diffusion Coefficient Reproducibility Across 3 T Scanners in a Breast Diffusion Phantom
- 27 August 2022
- journal article
- research article
- Published by Wiley in Journal of Magnetic Resonance Imaging
- Vol. 57 (3) , 812-823
Abstract
Background: To date, the accuracy and variability of diffusion‐weighted MRI (DW‐MRI) metrics have been reported in a limited number of scanner/protocol/coil combinations.Purpose: To evaluate the reproducibility of DW‐MRI estimates across multiple scanners and DW‐MRI protocols and to assess the effects of using an 8‐channel vs. 16‐channel breast coil in a breast phantom.Study Type: Prospective.Phantom: Breast phantom containing tubes of water and differing polyvinylpyrrolidone (PVP) concentrations with apparent diffusion coefficients (ADCs) matching breast tissue.Field Strength/Sequence: 3 T (three standard and one wide bore), three DW‐MRI single‐shot echo planar imaging protocols of varying acquired spatial resolution.Assessment: Accuracy of estimated ADCs was assessed using percent differences (PD) relative to nuclear magnetic resonance spectroscopy‐derived reference values. Coefficients of variation (CV) were calculated to determine variation across scanners. CVs based on the median standard deviation (CVM) were used to evaluate tube‐specific dispersion using 8‐ or 16‐channel coils at a single scanner. ADCs of PVP‐containing tubes were additionally normalized by the median water tube ADC to account for temperature effects.Statistical Tests: Two‐way repeated measures analysis of variance and post hoc tests were used to evaluate differences in ADC, CV, and CVM across scanners and protocols (α = 0.05).Results: ADCs were within 11% (interquartile range [IQR] 7%) of reference values and significantly improved to 2% (IQR 7%) after normalization to an internal water reference. Normalization significantly reduced interscanner variability of ADC estimates from 7% to 4%. DW‐MRI protocol did not affect ADC accuracy; however, the clinical and higher‐resolution clinical protocols resulted in the greatest (9%) and least (6%) interscanner variability, respectively. The 8‐ and 16‐channel receive coils yielded similar accuracy (PD: 12% vs. 16%) and precision (CVM: 2.7% vs. 2.9%).Data Conclusion: Normalization by an internal reference improved interscanner ADC reproducibility. High‐resolution protocols yielded comparably accurate and significantly less variable ADCs compared to a clinical standard protocol.Evidence Level: 2Technical Efficacy: Stage 1Keywords
Funding Information
- California Breast Cancer Research Program (IDEA Award 25IB‐0056)
- GE Healthcare
- National Cancer Institute (R37 CA249659)
This publication has 35 references indexed in Scilit:
- Apparent diffusion coefficient measurements in the differentiation between benign and malignant lesions: a systematic reviewInsights into Imaging, 2012
- Diffusion-weighted Imaging Improves the Diagnostic Accuracy of Conventional 3.0-T Breast MR ImagingRadiology, 2010
- Utility of diffusion‐weighted MRI in distinguishing benign and malignant hepatic lesionsJournal of Magnetic Resonance Imaging, 2010
- Quantitative diffusion weighted imaging for differentiation of benign and malignant breast lesions: The influence of the choice of b‐valuesJournal of Magnetic Resonance Imaging, 2010
- Efficient correction of inhomogeneous static magnetic field-induced distortion in Echo Planar ImagingNeuroImage, 2010
- Differential diagnosis of mammographically and clinically occult breast lesions on diffusion‐weighted MRIJournal of Magnetic Resonance Imaging, 2010
- Diffusion‐weighted imaging of breast cancer: Correlation of the apparent diffusion coefficient value with prognostic factorsJournal of Magnetic Resonance Imaging, 2009
- Influence of multichannel combination, parallel imaging and other reconstruction techniques on MRI noise characteristicsMagnetic Resonance Imaging, 2008
- Artifacts and pitfalls in diffusion MRIJournal of Magnetic Resonance Imaging, 2006
- Practical approaches to the evaluation of signal‐to‐noise ratio performance with parallel imaging: Application with cardiac imaging and a 32‐channel cardiac coilMagnetic Resonance in Medicine, 2005