Ventricular oversensing in 518 patients with implanted cardiac defibrillators: incidence, complications, and solutions
Authors:
Open Access
- 26 September 2007
- journal article
- research article
- Published by Oxford University Press (OUP) in EP Europace
- Vol. 9 (11) , 1041-1047
Abstract
The present study evaluates the incidence of various complications in implanted cardiac defibrillators (ICD) therapy due to ventricular oversensing (VO) and its complications. From June 1998 to May 2005, we retrospectively screened 518 patients (1085.6 patient years) for the occurrence of VO episodes (441 male, 77 female). The overall incidence was 7.3% (n = 38) with inappropriate shock deliveries accounting for 2.3% (n = 12). All VO episodes were caused by either T-wave oversensing (n = 10), myopotentials (n = 8), electrode failure (n = 5), interference with electromagnetic fields (n = 3), double-counting (n = 4), pacemaker interactions (n = 2), or others (n = 2). There were five life-threatening events due to inappropriate ICD reaction. In eight (22%) cases, ICD reprogramming was able to avoid further oversensing episodes (e.g. adaptation of sensitivity, T-wave suppression feature), 13 (35%) patients had to undergo invasive procedures (e.g. electrode replacing) to suppress VO, 16 (43%) were told to avoid the trigger situation, and one demanded to deactivate all ICD therapies because of inappropriate shock delivery. Our data demonstrate that VO is a rare complication, but might lead to life-threatening events. In most cases, VO episodes could be prevented by appropriate ICD reprogramming or avoidance of the initiating trigger.Keywords
This publication has 31 references indexed in Scilit:
- Amiodarone or an Implantable Cardioverter–Defibrillator for Congestive Heart FailureNew England Journal of Medicine, 2005
- Total Failure to Sense Ventricular Fibrillation with Inappropriate Defibrillator Sensitivity AdjustmentPacing and Clinical Electrophysiology, 2004
- Cardiac-Resynchronization Therapy with or without an Implantable Defibrillator in Advanced Chronic Heart FailureNew England Journal of Medicine, 2004
- Ventricular Oversensing:Pacing and Clinical Electrophysiology, 2003
- ACC/AHA/NASPE 2002 Guideline Update for Implantation of Cardiac Pacemakers and Antiarrhythmia Devices: Summary ArticleCirculation, 2002
- Prophylactic Implantation of a Defibrillator in Patients with Myocardial Infarction and Reduced Ejection FractionNew England Journal of Medicine, 2002
- Task Force on Sudden Cardiac Death of the European Society of CardiologyEuropean Heart Journal, 2001
- A Comparison of Antiarrhythmic-Drug Therapy with Implantable Defibrillators in Patients Resuscitated from Near-Fatal Ventricular ArrhythmiasNew England Journal of Medicine, 1997
- Unipolar Sensing Abnormalities: Incidence and Clinical Significance of Skeletal Muscle Interference and Under sensing in 228 PatientsPacing and Clinical Electrophysiology, 1982
- A comparison of unipolar and bipolar electrograms for cardiac pacemaker sensing.Circulation, 1977