Validation of Air as an Equilibration Medium in Gastric Tonometry: An in Vitro Evaluation of Two Techniques for Measuring Air Pco 2

Authors:

Abstract
This laboratory-based bench study was undertaken to evaluate the accuracy and equilibration characteristics of air and saline respectively as CO 2 equilibrating media in the silicone balloon of a gastric tonometer and to compare two methods of measuring air Pco 2 . Two gastric tonometers were suspended in a bath containing 0.9% saline maintained at 37°C. Certified calibration gases at three different CO 2 concentrations were bubbled into the bath. When the bath Pco 2 measurement was stable the tonometers were primed with 5 ml of air and 2.5 ml normal saline respectively and allowed to equilibrate for 30 and 90 minutes. Following equilibration, samples were aspirated and analysed in duplicate in a blood gas analyser. Bias and precision were calculated from the measured and expected Pco 2 values. A consistent negative bias (21–23%) was seen with air at all three CO 2 concentrations at 30 and 90 minutes with a coefficient of variation between 2.7 and 3.3%. Imprecise data were obtained with saline at different levels of CO 2 . A similar experimental set-up was used to compare air Pco 2 measurement by a blood gas analyser and an infra-red analyser (Tonocap). Similar bias was obtained with the blood gas analyser with respect to air Pco 2 measurement as in experiment 1. The infra-red analyser measurement was highly precise with negligible bias. Air appears to be a better CO 2 equilibration medium during bench testing of tonometry producing a systematic negative offset and requiring a uniform correction factor of 1.25. This correction factor is independent of equilibration time and equilibrating CO 2 concentration. The use of the infra-red analyser eliminates any bias in the measurement of air Pco 2 and obviates the need for a correction factor.

This publication has 17 references indexed in Scilit: