Glow discharge electron impact ionization source for miniature mass spectrometers
- 18 April 2007
- journal article
- research article
- Published by Wiley in Journal of Mass Spectrometry
- Vol. 42 (5) , 675-680
Abstract
A glow discharge electron impact ionization (GDEI) source was developed for operation using air as the support gas. An alternative to the use of thermoemission from a resistively heated filament electron source for miniature mass spectrometers, the GDEI source is shown to have advantages of long lifetime under high‐pressure operation and low power consumption. The GDEI source was characterized using our laboratory's handheld mass spectrometer, the Mini 10. The effects of the discharge voltage and pressure were investigated. Design considerations are illustrated with calculations. Performance is demonstrated in a set of experimental tests. The results show that the low power requirements, mechanical ruggedness, and quality of the data produced using the small glow discharge ion source make it well‐suited for use with a portable handheld mass spectrometer. Copyright © 2007 John Wiley & Sons, Ltd.Keywords
This publication has 21 references indexed in Scilit:
- Handheld Rectilinear Ion Trap Mass SpectrometerAnalytical Chemistry, 2006
- Microfabricated Quadrupole Ion Trap for Mass Spectrometer ApplicationsPhysical Review Letters, 2006
- A Quadrupole Ion Trap with Cylindrical Geometry Operated in the Mass-Selective Instability ModeAnalytical Chemistry, 1998
- Miniature, high-resolution, quadrupole mass-spectrometer arrayReview of Scientific Instruments, 1997
- High-pressure effects in miniature arrays of quadrupole analyzers for residual gas analysis from 10−9 to 10−2 TorrJournal of Vacuum Science & Technology A, 1996
- Ion trap mass spectrometry of externally generated ionsAnalytical Chemistry, 1994
- Role of gas dynamics in negative ion formation in an atmospheric sampling glow discharge ionization sourceAnalytical Chemistry, 1993
- Coupling of an atmospheric-samling ion source with an ion-trap mass spectrometerAnalytica Chimica Acta, 1989
- Hollow cathode plume as an atomic emission source for elemental analysis of metal alloysAnalytical Chemistry, 1986
- Scientific Foundations of Vacuum TechniqueAmerican Journal of Physics, 1962