Inter- and Intraobserver Reliability of Two-dimensional CT Scan for Total Knee Arthroplasty Component Malrotation
Authors:
- Ryan HessRHRyan Hess
- Curtis HartmanCHCurtis Hartman
- Lynette SmithLSLynette Smith
- Kevin L. GarvinKevin L. Garvin
- 1 January 2014
- journal article
- research article
- Published by Wolters Kluwer Health in Clinical Orthopaedics and Related Research
- Vol. 472 (1) , 212-217
Abstract
F CT scans for this use has not been well documented. Questions/purposes The objective of this study was to determine the inter- and intraobserver reliability and the repeatability of the measurement of TKA component rotation using two-dimensional CT scans. Methods Fifty-two CT scans of TKAs being evaluated for revision surgery were measured by three different physicians. An orthopaedic resident and attending measured the same scans twice (more than 2 weeks apart) and a musculoskeletal radiologist measured them once. To assess interobserver reliability, intraclass correlation coefficients (ICCs) with two-way mixed-effects analysis of variance models as well as 95% confidence intervals for each were done. The repeatability coefficient was calculated as well, which is defined as the difference in measurements that include 95% of the values. This indicates the magnitude of variability among measurements in the same scale, which in this study is degrees. Results The interobserver ICC measurement for the femoral component was 0.386 (poor), and it was 0.670 (good) for the tibial component. The interobserver ICC for the combined rotation measurements was 0.617 (good). The intraobserver ICC for the femoral component was 0.606 (good), and it was 0.809 (very good) for the tibial component. The intraobserver ICC for combined rotation was 0.751 (good). The intraobserver repeatability coefficient for the femoral component was 0.49°, 10.64° for the tibial component, and 12.29° for combined rotation. Conclusions In this study, the inter- and intraobserver reliability, and the repeatability, of TKA component rotation were variable. This raises concern about whether CT scan is diagnostic in the assessment of component malrotation after TKA. Level of Evidence Level IV, diagnostic study. See Guidelines for Authors for a complete description of levels of evidence....Keywords
This publication has 22 references indexed in Scilit:
- Internal Rotation of the Tibial Component is Frequent in Stiff Total Knee ArthroplastyClinical Orthopaedics and Related Research, 2011
- The position and orientation of total knee replacement componentsThe Journal of Bone and Joint Surgery. British volume, 2011
- Early Revision for Component Malrotation in Total Knee ArthroplastyClinical Orthopaedics and Related Research, 2007
- Axial Radiography of the Distal Femur to Assess Rotational Alignment in Total Knee ArthroplastyClinical Orthopaedics and Related Research, 2005
- Revision of the stiff total knee arthroplastyThe Journal of Arthroplasty, 2002
- The accuracy of computed tomography for determining femoral and tibial total knee arthroplasty component rotationThe Journal of Arthroplasty, 2000
- Effect of rotational malposition of the femoral component on knee stability kinematics after total knee arthroplastyThe Journal of Arthroplasty, 1995
- The Patellofemoral Joint in Total Condylar Knee ArthroplastyClinical Orthopaedics and Related Research, 1986
- STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENTThe Lancet, 1986
- Total Knee Dislocation Due to Rotatory Malalignment of Tibial ComponentClinical Orthopaedics and Related Research, 1980