Sensitivity, robustness, and reproducibility of U-shaped delamination test for evaluation of candidate ultra-high molecular weight polyethylene materials for joint replacements
Authors:
- 3 June 2024
- journal article
- research article
- Published by SAGE Publications in Proceedings of the Institution of Mechanical Engineers, Part H: Journal of Engineering in Medicine
- Vol. 238 (7) , 764-773
Abstract
The delamination of ultra-high molecular weight polyethylene (UHMWPE) in artificial joints is a major cause limiting the long-term clinical results of arthroplasty. However, the conventional test method using simple reciprocation to evaluate the delamination resistance of UHMWPE materials has insufficient detection sensitivity. To reproduce delamination, the unconformity contact must be maintained throughout the test so that the maximum stress is generated below the surface. Therefore, a test method that applies a U-shaped motion comprising two long-linear and one short linear sliding motion was developed. The sensitivity, robustness, and reproducibility of the U-shaped delamination test were investigated and compared with the traditional test method. The traditional test method could reproduce delamination only in materials that had degraded considerably, whereas the U-shaped delamination test could reproduce delamination in a wide range of materials, demonstrating its superior sensitivity. Additionally, using a higher load helped accelerate the test without affecting the test results. The optimal length of the short linear sliding motion was confirmed to be 1 mm. Finally, the inter-laboratory reproducibility of the U-shaped delamination test was confirmed using the round-robin test. The U-shaped delamination test demonstrates high sensitivity, robustness, and reproducibility and contributes to the selection and development of UHMWPE materials and artificial joints with a lower risk of delamination.Keywords
Funding Information
- Japan Society for the Promotion of Science (JP22K09437)
- Japan Agency for Medical Research and Development (JP20mk0102116)
This publication has 31 references indexed in Scilit:
- The efficacy of antibiotic‐eluting material in a two‐stage model of periprosthetic joint infectionJournal of Orthopaedic Research, 2023
- In vitro estimation of reduction in strength and wear resistance of UHMWPE for joint prostheses due to lipid‐induced degradationJournal of Biomedical Materials Research Part B: Applied Biomaterials, 2020
- Evaluating the durability of UHMWPE biomaterials used for articulating surfaces of joint arthroplasty using delamination testsJournal of Biomedical Materials Research Part B: Applied Biomaterials, 2018
- Wear, delamination, and fatigue resistance of melt‐annealed highly crosslinked UHMWPE cruciate‐retaining knee inserts under activities of daily livingJournal of Orthopaedic Research, 2010
- The effect of oxidative aging on the wear performance of highly crosslinked polyethylene knee inserts under conditions of severe malalignmentJournal of Orthopaedic Research, 2008
- A Highly Crosslinked UHMWPE for CR and PS Total Knee ArthroplastiesThe Journal of Arthroplasty, 2008
- Knee-simulator testing of conventional and cross-linked polyethylene tibial insertsThe Journal of Arthroplasty, 2004
- Prevention of fatigue cracks in ultrahigh molecular weight polyethylene joint components by the addition of vitamin EJournal of Biomedical Materials Research, 1999
- Retrieval analysis of total knee prostheses: A method and its application to 48 total condylar prosthesesJournal of Biomedical Materials Research, 1983
- Reactions of the articular capsule to wear products of artificial joint prosthesesJournal of Biomedical Materials Research, 1977