Completely pulsatile high flow circulatory support with a constant-speed centrifugal blood pump: mechanisms and early clinical observations
Authors:
- 11 April 2007
- journal article
- Published by Springer Nature in General Thoracic and Cardiovascular Surgery
- Vol. 55 (4) , 158-162
Abstract
Various types of rotary blood pumps (axial flow, centrifugal) have been introduced into clinical use recently. These pumps have different pressure–flow characteristics, and some investigators have noted that a limited pump flow rate and less pulsatility are the problems with the axial flow devices. A new implantable centrifugal blood pump was developed that has an extremely flat pressure–flow curve and is able to produce a significantly high pump flow rate of 20 l/min at a low pressure of 10–30 mmHg. When the pressure difference between the left ventricle and aorta decreases during systole, an instant high peak flow is achieved, which results in a higher peak pressure in the aorta (systolic pressure). During the diastolic phase, the left ventricle–aorta pressure difference increases to maximum, and the pump flow rate decreases to minimum. Thus, the pump flow rate becomes completely pulsatile, and the high peak flow provides a higher mean pump flow rate. This pump was applied to two end-stage heart failure patients (dilated cardiomyopathy, New York Heart Association (NYHA) class IV, inotrope-dependent). The pump was observed to provide completely pulsatile high flow assistance of 6–9 l/min with a constant pump speed. Both patients are currently in NYHA class I after 1 year on the device with no major adverse events. The new centrifugal blood pump provides completely pulsatile high-flow circulatory support with a constant pump speed, which solves the current clinical problems with rotary blood pumps.Keywords
This publication has 5 references indexed in Scilit:
- Use of the Jarvik 2000 Left Ventricular Assist System as a Bridge to Heart Transplantation or as Destination Therapy for Patients With Chronic Heart FailureAnnals of Surgery, 2003
- EVAHEART™: An implantable centrifugal blood pump for long-term circulatory supportThe Japanese Journal of Thoracic and Cardiovascular Surgery, 2002
- Pulsatile Flow in Patients With a Novel Nonpulsatile Implantable Ventricular Assist DeviceCirculation, 2000
- Rotary Blood Pump Flow Spontaneously Increases During Exercise Under Constant Pump Speed: Results of a Chronic StudyArtificial Organs, 1999