Flame atomic absorption spectrometric determination of cadmium in biological samples using a preconcentration flow system with an activated carbon column and dithizone as a chelating agent
Authors:
- Yaneira Petit de PeñaYPYaneira Petit de Peña
- Mercedes GallegoMGMercedes Gallego
- Miguel ValcárcelMVMiguel Valcárcel
- 1 January 1994
- journal article
- research article
- Published by Royal Society of Chemistry (RSC) in Journal of Analytical Atomic Spectrometry
- Vol. 9 (6) , 691-696
Abstract
A combined flow atomic absorption spectrometric system was used to develop an efficient on-line preconcentration-solvent elution procedure for the determination of trace amounts of cadmium. The metal was preconcentrated as the dithizonate on a mini-column packed with activated carbon and eluted with isobutyl methyl ketone (4-methylpentan-2-one) into the nebulizer. A preconcentration factor of 40–130, equivalent to 3–12 ml of sample, was achieved by using a time-based technique. The detection limit obtained (3σ) ranged between 0.3 and 1.3 ng ml–1 and the relative standard deviation from 1.3 to 2.4% for a sample volume of 12 and 3 ml, respectively. The results obtained for the determination of cadmium in reference materials testify to the applicability of the proposed procedure to the analysis of biological materials.Keywords
This publication has 27 references indexed in Scilit:
- Determination of molybdenum by flame and electrothermal atomisation atomic absorption spectrometry after complexation and sorption on activated carbonJournal of Analytical Atomic Spectrometry, 1990
- Enrichment of trace metals in water on activated carbonThe Analyst, 1990
- Novel preconcentration technique for the determination of trace elements in fine chemicalsAnalytical Proceedings including Analytical Communications, 1989
- Pre-concentration of trace metals from acidic sample solutions on a thin layer of activated carbon using dithizone as a chelating agentThe Analyst, 1988
- On-line trace metal enrichment and matrix isolation in atomic absorption spectrometry by a column containing immobilized 8-quinolinol in a flow-injection systemAnalytica Chimica Acta, 1984
- An efficient flow-injection system with on-line ion-exchange preconcentration for the determination of trace amounts of heavy metals by atomic absorption spectrometryAnalytica Chimica Acta, 1984
- Combination of flow injection analysis with flame atomic-absorption spectrophotometry: determination of trace amounts of heavy metals in polluted seawaterThe Analyst, 1983
- Preconcentration of traces of silver and bismuth from cobalt and nickel and their nitrates, with activated carbonTalanta, 1979
- Enrichment of trace metals in water by adsorption on activated carbonAnalytical Chemistry, 1977
- Trace enrichment with activated carbon and determination of trace metals in high-purity zinc and zinc(II) nitrateTalanta, 1977