A multi-slot surface coil for MRI of dual-rat imaging at 4 T
Authors:
- S E SolisSSS E Solis
- R WangRWR Wang
- D TomasiDTD Tomasi
- A O RodriguezARA O Rodriguez
- 23 May 2011
- journal article
- Published by IOP Publishing in Physics in Medicine & Biology
- Vol. 56 (12) , 3551-3561
Abstract
A slotted surface coil inspired by the hole-and-slot cavity magnetron was developed for magnetic resonance imaging of obese rats at 4 T. Full-wave analysis of the magnetic field was carried out at 170 MHz for both the slotted and circular-shaped coils. The noise figure values of two coils were investigated via the numerical calculation of the quality factors. Fat simulated phantoms to mimic overweight rats were included in the analysis with weights ranging from 300 to 900 g. The noise figures were 1.2 dB for the slotted coil and 2.4 dB for the circular coil when loaded with 600 g of simulated phantom. A slotted surface coil with eight circular slots and a circular coil with similar dimensions were built and operated in the transceiver mode, and their performances were experimentally compared. The imaging tests in phantoms demonstrated that the slotted surface coil has a deeper RF-sensitivity and better field uniformity than the single-loop RF-coil. High quality images of two overweight Zucker rats were acquired simultaneously with the slotted surface coil using standard spin-echo pulse sequences. Experimental results showed that the slotted surface coil outperformed the circular coil for imaging considerably overweight rats. Thus, the slotted surface coil can be a good tool for MRI experiments in rats on a human whole-body 4 T scanner.Keywords
This publication has 5 references indexed in Scilit:
- A four-channel hole-slotted phased array at 7 TeslaConcepts in Magnetic Resonance Part B, 2010
- Noise figure limits for circular loop MR coilsMagnetic Resonance in Medicine, 2009
- Radio frequency coil technology for small‐animal MRINMR in Biomedicine, 2007
- Magnetron Surface Coil for Brain MR ImagingArchives of Medical Research, 2006
- The dielectric properties of biological tissues: II. Measurements in the frequency range 10 Hz to 20 GHzPhysics in Medicine & Biology, 1996