Evaluating the Effect of Flow and Interface Type on Pressures Delivered With Bubble CPAP in a Simulated Model
Authors:
Open Access
- 3 November 2015
- journal article
- research article
- Published by SAGE Publications in Respiratory Care
- Vol. 61 (3) , 333-339
Abstract
BACKGROUND: Bubble CPAP, used for spontaneously breathing infants to avoid intubation or postextubation support, can be delivered with different interface types. This study compared the effect that interfaces had on CPAP delivery. We hypothesized that there would be no difference between set and measured levels between interface types. METHODS: A validated preterm infant nasal airway model was attached to the ASL 5000 breathing simulator. The simulator was programmed to deliver active breathing of a surfactant-deficient premature infant with breathing frequency at 70 breaths/min inspiratory time of 0.30 s, resistance of 150 cm H 2 O/L/s, compliance of 0.5 mL/cm H 2 O, tidal volume of 5 mL, and esophageal pressure of −10 cm H 2 O. Nasal CPAP prongs, size 4030, newborn and infant RAM cannulas were connected to a nasal airway model and a bubble CPAP system. CPAP levels were set at 4, 5, 6, 7, 8, and 9 cm H 2 O with flows of 6, 8, and 10 L/min each. Measurements were recorded after 1 min of stabilization. The analysis was performed using SAS 9.4. The Kolmogorov-Smirnov test assessed normality of the data. The Friedman test was used to compare non-normally distributed repeated measures. The Wilcoxon signed-rank test was used to conduct post hoc analysis. All tests were 2-sided, and P values of <.05 were considered as indicating significant differences unless otherwise indicated. RESULTS: At lower set CPAP levels, 4–6 cm H 2 O, measured CPAP dropped precipitously with the nasal prongs with the highest flow setting. At higher CPAP levels, 7–9 cm H 2 O measured CPAP concomitantly increased as the flow setting increased. Statistically significant differences in set and measured CPAP occurred for all devices across all CPAP levels, with the measured CPAP less than set for all conditions, P < .001. CONCLUSIONS: Set flow had a profound effect on measured CPAP. The concomitant drop in measured pressure with high and low flows could be attributed to increased resistance to spontaneous breathing or insufficient flow to meet inspiratory demand. Clinicians should be aware of the effect that the interface and flow have on CPAP delivery.Keywords
This publication has 14 references indexed in Scilit:
- Evaluation of a Nasal Cannula in Noninvasive Ventilation Using a Lung SimulatorRespiratory Care, 2014
- Volume Oscillations Delivered to a Lung Model Using 4 Different Bubble CPAP SystemsRespiratory Care, 2014
- An Integrative Review of Skin Breakdown in the Preterm Infant Associated with Nasal Continuous Positive Airway PressureJournal of Obstetric, Gynecologic, & Neonatal Nursing, 2013
- Effects of Condensate in the Exhalation Limb of Neonatal Circuits on Airway Pressure During Bubble CPAPRespiratory Care, 2013
- Randomized Trial Comparing 3 Approaches to the Initial Respiratory Management of Preterm NeonatesPediatrics, 2011
- Early CPAP versus Surfactant in Extremely Preterm InfantsNew England Journal of Medicine, 2010
- Work of Breathing During Nasal Continuous Positive Airway Pressure in Preterm Infants: A Comparison of Bubble vs Variable-Flow DevicesJournal of Perinatology, 2005
- Airway Obstruction in Two Extremely Low Birthweight Infants Treated with Oxygen CannulasJournal of Perinatology, 2003
- Use of Oxygen Cannulas in Extremely Low Birthweight Infants is Associated with Mucosal Trauma and Bleeding, and Possibly with Coagulase-negative Staphylococcal SepsisJournal of Perinatology, 2003
- Computerised determination of spontaneous inspiratory and expiratory times in premature neonates during intermittent positive pressure ventilation. II: Results from 20 babies.Archives of Disease in Childhood - Fetal and Neonatal Edition, 1994