International normalized ratio (INR) testing in Europe: between-laboratory comparability of test results obtained by Quick and Owren reagents
Authors:
- 12 April 2018
- journal article
- research article
- Published by Walter de Gruyter GmbH in Clinical Chemistry and Laboratory Medicine (CCLM)
- Vol. 56 (10) , 1698-1703
Abstract
Background: This study was designed to obtain an overview of the analytical quality of the prothrombin time, reported as international normalized ratio (INR) and to assess the variation of INR results between European laboratories, the difference between Quick-type and Owren-type methods and the effect of using local INR calibration or not. In addition, we assessed the variation in INR results obtained for a single donation in comparison with a pool of several plasmas. Methods: A set of four different lyophilized plasma samples were distributed via national EQA organizations to participating laboratories for INR measurement. Results: Between-laboratory variation was lower in the Owren group than in the Quick group (on average: 6.7% vs. 8.1%, respectively). Differences in the mean INR value between the Owren and Quick group were relatively small (0.5) were observed between different reagents. No systematic significant differences in the between-laboratory variation for a single-plasma sample and a pooled plasma sample were observed. Conclusions: The comparability for laboratories using local calibration of their thromboplastin reagent is better than for laboratories not performing local calibration. Implementing local calibration is strongly recommended for the measurement of INR.Keywords
This publication has 16 references indexed in Scilit:
- Local Verification and Assignment of Mean Normal Prothrombin Time and International Sensitivity Index Values across Various Instruments: Recent Experience and Outcome from North AmericaSeminars in Thrombosis and Hemostasis, 2013
- Self-monitoring of oral anticoagulation: systematic review and meta-analysis of individual patient dataThe Lancet, 2012
- A sensitivity comparison of the Quick and Owren prothrombin time methods in oral anticoagulant therapyHematology Reports, 2009
- Systematic Overview of Warfarin and Its Drug and Food InteractionsArchives of internal medicine (1960), 2005
- Poor Agreement among Prothrombin Time International Normalized Ratio Methods: Comparison of Seven Commercial ReagentsClinical Chemistry, 2005
- Guidelines on preparation, certification, and use of certified plasmas for ISI calibration and INR determinationJournal of Thrombosis and Haemostasis, 2004
- Local INR calibration of the Owren type prothrombin assay greatly improves the intra- and interlaboratory variationThrombosis and Haemostasis, 2004
- Comparison of Quick and Owren Prothrombin Time with Regard to the Harmonisation of the International Normalised Ratio (INR) SystemClinical Chemistry and Laboratory Medicine (CCLM), 2002
- The Control of Dicumarol Therapy and the Quantitative Determination of Prothrombin and ProconvertinScandinavian Journal of Clinical and Laboratory Investigation, 1951
- A STUDY OF THE COAGULATION DEFECT IN HEMOPHILIA AND IN JAUNDICE.*The American Journal of the Medical Sciences, 1935