Repeatability of Spirometry in 18,000 Adult Patients
Authors:
- 15 January 2004
- journal article
- Published by Oxford University Press (OUP) in American Journal of Respiratory and Critical Care Medicine
- Vol. 169 (2) , 235-238
Abstract
The objective of this study was to determine the limits for repeatability of FEV1, FVC, and PEF during spirometry test sessions in adult outpatients. A retrospective chart review of 18,000 consecutive patients, aged 20 to 90 years, referred to a large outpatient pulmonary function laboratory for testing was performed. Measurements included the differences between the highest and second-highest FVC (dFVC), FEV1 (dFEV1), and PEF (dPEF), from prebronchodilator spirometry, and anthropometric factors. Ninety percent of the patients were able to reproduce FEV1 within 120 ml (6.1%), FVC within 150 ml (5.3%), and PEF within 0.80 L (12%). Patient characteristics, such as sex, age, height, smoking status, and FEV1 (% predicted), had very little effect on repeatability, explaining only 2 to 4% of the variation in repeatability (expressed in milliliters). We conclude that the ability of patients to meet or exceed spirometry repeatability goals does not depend on patient characteristics when testing is performed by experienced personnel. The current American Thoracic Society repeatability goal of 200 ml for FEV1 and FVC may be too lenient.Keywords
This publication has 17 references indexed in Scilit:
- Quality of Spirometry Test Performance in Children and Adolescents*Chest, 2000
- Quality Control of Spirometry Testing in the Registry for Patients With Severe α1-Antitrypsin DeficiencyChest, 1997
- Standardization of Spirometry, 1994 Update. American Thoracic SocietyAmerican Journal of Respiratory and Critical Care Medicine, 1995
- Predictors of spirometric test failure: a comparison of the 1983 and 1993 acceptability criteria from the European Community for Coal and Steel.Occupational and Environmental Medicine, 1995
- Cognitive Function and Spirometry Performance in the ElderlyAmerican Review of Respiratory Disease, 1993
- Lung volumes and forced ventilatory flowsEuropean Respiratory Journal, 1993
- Spirometric Lung Function: Distribution and Determinants of Test Failure in a Young Adult PopulationAmerican Review of Respiratory Disease, 1992
- Lung Function Testing: Selection of Reference Values and Interpretative StrategiesAmerican Review of Respiratory Disease, 1991
- Acceptability and Reproducibility Criteria of the American Thoracic Society as Observed in a Sample of the General PopulationAmerican Review of Respiratory Disease, 1991
- Standardization of Spirometry—1987 UpdateAmerican Review of Respiratory Disease, 1987