A design of an ECR ion source for radioactive ion beams for ISAC on-line facility at TRIUMF
Authors:
- 1 February 2002
- journal article
- conference paper
- Published by AIP Publishing in Review of Scientific Instruments
- Vol. 73 (2) , 792-794
Abstract
A 2.45 GHz ECR ion source for the ISAC on-line operation is being tested at the TRIUMF ion source test stand prior to its installation in the radioactive ion beam production target module. Steps were taken at the design stage to ensure short transient time and high ionization efficiency for single charged ions. Further considerations were given to radiation hardness, reliability, and remote handling capabilities. Transient time was measured using an ultrafast piezoelectric valve. Efficiencies were measured using calibrated gas leaks of stable and radioactive isotopes such as This article describes the overall characteristics of the source.
Keywords
This publication has 5 references indexed in Scilit:
- Efficiency and transient time studies of an electron cyclotron resonance ion source for radioactive ion beam production at ISAC/TRIUMFReview of Scientific Instruments, 2000
- Commissioning and initial operation of a radioactive beam ion source at ISACReview of Scientific Instruments, 2000
- Design of an electron cyclotron resonance ion source for the isotope separator and accelerator at TRIUMFReview of Scientific Instruments, 2000
- Design of a parallel-plate energy spread analyzerReview of Scientific Instruments, 1998
- An electron cyclotron resonance source for radioactive beryllium ion beam productionReview of Scientific Instruments, 1998