Left main coronary artery bidirectional dissection caused by ejection of guideliner catheter from the guiding catheter
Authors:
- 25 March 2013
- journal article
- case report
- Published by Wiley in Catheterization and Cardiovascular Interventions
- Vol. 82 (3) , E215-20
Abstract
The “Guideliner,” a child catheter, aims to deliver stents in tortuous coronary lesions by deep engagement and providing better support. Coronary artery spiral dissection, which may be caused by deep‐seated coronary catheters or noncoaxial engagement is one of the most seen percutaneous coronary intervention‐related complications. In this case, Guideliner‐related left main dissection occurred after successful revascularization of a chronic total occlusion at the left anterior descending artery and the left circumflex, which has not been previously reported. The Guideliner catheter is efficient in treating complex coronary lesions, but one should be aware of its related complications.Keywords
This publication has 6 references indexed in Scilit:
- Guideliner® catheter—Friend or foe?Catheterization and Cardiovascular Interventions, 2011
- Transfemoral use of the guideLiner catheter in complex coronary and bypass graft interventionsCatheterization and Cardiovascular Interventions, 2011
- The use of a guide catheter extension system as an aid during transradial percutaneous coronary intervention of coronary artery bypass graftsCatheterization and Cardiovascular Interventions, 2011
- Use of the GuideLiner catheter in facilitating coronary and graft interventionCardiovascular Revascularization Medicine, 2011
- Distal stent delivery with guideliner catheter: First in man experienceCatheterization and Cardiovascular Interventions, 2010
- Impact of complete revascularization with percutaneous coronary intervention on survival in patients with at least one chronic total occlusionEuropean Heart Journal, 2008