Measurement of Adult Human Brain Responses to Breath Holding by Multi-Distance Hyperspectral Near-Infrared Spectroscopy

Authors:
  • ZGZahida Guerouah
    Zahida Guerouah
  • SLSteve Lin
    Steve Lin
  • VTVladislav Toronov
    Vladislav Toronov

Abstract
A major limitation of near-infrared spectroscopy (NIRS) is its high sensitivity to the scalp and low sensitivity to the brain of adult humans. In the present work we use multi-distance hyperspectral NIRS (hNIRS) to investigate the optimal source-detector distances, range of wavelengths, and analysis techniques to separate cerebral responses to 30-s breath holds (BHs) from the responses in the superficial tissue layer in healthy adult humans. We observed significant responses to BHs in the scalp hemodynamics. Cerebral responses to BHs were detected in the cytochrome C oxidase redox (rCCO) at 4 cm without using data from the short-distance channel. Using the data from the 1 cm channel in the two-layer regression algorithm showed that hemodynamic and rCCO responses also occurred at 3cm. We found that the waveband 700-900 nm was optimal for the detection of cerebral responses to BHs in adults.