High-Resolution Infrared Thermal Imaging of the Esophagus During Atrial Fibrillation Ablation as a Predictor of Endoscopically Detected Thermal Lesions
Authors:
- 1 November 2018
- journal article
- research article
- Published by Wolters Kluwer Health in Circulation: Arrhythmia and Electrophysiology
- Vol. 11 (11) , e006681
Abstract
Background: Endoscopically detected thermal esophageal lesions (EDEL) after ablation of atrial fibrillation may be precursors of atrioesophageal fistula and esophageal luminal temperature monitoring has previously failed to decrease thermal damage. Methods: Sixty-three patients undergoing their first pulmonary vein isolation using radiofrequency point-by-point catheter ablation were prospectively included in the HEAT-AF study (High-Resolution Esophageal Assessment of Esophageal Temperature During Atrial Fibrillation Ablation) and esophageal temperatures were continuously monitored using a novel infrared thermography system (IRTS). Peak esophageal temperature (Tpeak) was correlated to postablation endoscopy results characterizing patients as EDEL positive or negative. Results: Twelve patients had EDEL (19%). Comparing EDEL positive to negative patients, Tpeak was significantly higher (56.3±4.6°C versus 45.7±5.5°C, P P =0.0008) of EDEL and yielded an odds ratio of 1.52; 95% CI, (1.24–2.05). Receiver operator curve analysis demonstrated Tpeak as a highly accurate binary classifier with an area under the curve of 93%. Conclusions: For the first time esophageal temperature monitoring using a high resolution, high-fidelity IRTS allowed accurate prediction of postablation EDEL suggesting that Tpeak alone is an excellent binary classifier of patients at risk of EDEL. The logistic regression model and associated receiver operator curve will aid in the selection of optimal temperature thresholds in future prospective studies.Keywords
This publication has 22 references indexed in Scilit:
- Multi‐Sensor Esophageal Temperature Probe Used During Radiofrequency Ablation for Atrial Fibrillation is Associated with Increased Intraluminal Temperature Detection and Increased Risk of Esophageal Injury Compared to Single‐Sensor ProbeJournal of Cardiovascular Electrophysiology, 2013
- Direct Measurement of the Lethal Isotherm for Radiofrequency Ablation of Myocardial TissueCirculation: Arrhythmia and Electrophysiology, 2011
- Utility of Esophageal Temperature Monitoring During Pulmonary Vein Isolation for Atrial Fibrillation Using Duty-Cycled Phased Radiofrequency AblationJournal of Cardiovascular Electrophysiology, 2010
- Esophageal Capsule Endoscopy After Radiofrequency Catheter Ablation for Atrial FibrillationCirculation: Arrhythmia and Electrophysiology, 2009
- Canine Model of Esophageal Injury and Atrial-Esophageal Fistula After Applications of Forward-Firing High-Intensity Focused Ultrasound and Side-Firing Unfocused Ultrasound in the Left Atrium and Inside the Pulmonary VeinCirculation: Arrhythmia and Electrophysiology, 2009
- Esophagus histological analysis after hyperthermia-induced injury: Implications for cardiac ablationInternational Journal of Hyperthermia, 2009
- Esophageal Injury and Temperature Monitoring During Atrial Fibrillation AblationCirculation: Arrhythmia and Electrophysiology, 2008
- Computed Tomographic Analysis of the Anatomy of the Left Atrium and the EsophagusCirculation, 2004
- Left Atrial–Esophageal Fistula Following Radiofrequency Catheter Ablation of Atrial FibrillationJournal of Cardiovascular Electrophysiology, 2004
- Atrio-Esophageal Fistula as a Complication of Percutaneous Transcatheter Ablation of Atrial FibrillationCirculation, 2004