Insertion torque, resonance frequency, and removal torque analysis of microimplants
Open Access
- 24 August 2016
- journal article
- Published by Wiley in The Kaohsiung Journal of Medical Sciences
- Vol. 32 (9) , 469-474
Abstract
This study aimed to compare the insertion torque (IT), resonance frequency (RF), and removal torque (RT) among three microimplant brands. Thirty microimplants of the three brands were used as follows: Type A (titanium alloy, 1.5‐mm × 8‐mm), Type B (stainless steel, 1.5‐mm × 8‐mm), and Type C (titanium alloy, 1.5‐mm × 9‐mm). A synthetic bone with a 2‐mm cortical bone and bone marrow was used. Each microimplant was inserted into the synthetic bone, without predrilling, to a 7 mm depth. The IT, RF, and RT were measured in both vertical and horizontal directions. One‐way analysis of variance and Spearman's rank correlation coefficient tests were used for intergroup and intragroup comparisons, respectively. In the vertical test, the ITs of Type C (7.8 Ncm) and Type B (7.5 Ncm) were significantly higher than that of Type A (4.4 Ncm). The RFs of Type C (11.5 kHz) and Type A (10.2 kHz) were significantly higher than that of Type B (7.5 kHz). Type C (7.4 Ncm) and Type B (7.3 Ncm) had significantly higher RTs than did Type A (4.1 Ncm). In the horizontal test, both the ITs and RTs were significantly higher for Type C, compared with Type A. No significant differences were found among the groups, and the study hypothesis was accepted. Type A had the lowest inner/outer diameter ratio and widest apical facing angle, engendering the lowest IT and highest RF values. However, no significant correlations in the IT, RF, and RT were observed among the three groups.Keywords
Funding Information
- Kaohsiung Medical University Chung-Ho Memorial Hospital (KMUH99‐9M30)
- Kaohsiung Medical University Hospital, Taiwan (KMUH99-9M30)
This publication has 15 references indexed in Scilit:
- Insertion torque and success of orthodontic mini-implants: A systematic reviewAmerican Journal of Orthodontics and Dentofacial Orthopedics, 2012
- Impact of bone quality, implant type, and implantation site preparation on insertion torques of mini-implants used for orthodontic anchorageInternational Journal of Oral and Maxillofacial Surgery, 2011
- In vitro evaluation of insertion and removal torques of orthodontic mini-implantsInternational Journal of Oral and Maxillofacial Surgery, 2011
- Clinical study on the insertion torque and wireless resonance frequency analysis in the assessment of torque capacity and stability of self‐tapping dental implantsJournal of Oral Rehabilitation, 2009
- Three-dimensional evaluation of interradicular spaces and cortical bone thickness for the placement and initial stability of microimplants in adultsAmerican Journal of Orthodontics and Dentofacial Orthopedics, 2009
- Factors Associated With the Stability of Mini-Implants for Orthodontic Anchorage: A Study of 414 Samples in TaiwanJournal of Oral and Maxillofacial Surgery, 2009
- A torque removal study on the primary stability of orthodontic titanium screw mini-implants in the cortical bone of dog femursInternational Journal of Oral and Maxillofacial Surgery, 2008
- Soft Tissue Thickness for Placement of an Orthodontic Miniscrew Using an Ultrasonic DeviceThe Angle Orthodontist, 2008
- Relevance of resonance frequency analysis to evaluate dental implant stability: simulation and histomorphometrical animal experimentsClinical Oral Implants Research, 2007
- The application of mini-implants for orthodontic anchorageInternational Journal of Oral and Maxillofacial Surgery, 2006