Optimization of Plasma Treatment, Manipulative Variables and Coating Composition for the Controlled Filling and Coating of a Microstructured Reservoir Stent
- 1 March 2009
- journal article
- Published by ASME International in Journal of Medical Devices
- Vol. 3 (1) , 011005
Abstract
The object of the study was to fill and coat the microcavities of a drug eluting stent using a batch dipping process. 316L coronary stents, which were coated with a layer of were used as substrates. The stents’ surface was dimpled with microcavities separated by distances of depending on location. The experiment consisted of (1) optimizing the procedures to fill the microcavities with a solution of therapeutic agent and (2) covering the filled microcavities with a protective “lid” that shielded the solution during stent insertion in the arteries and then controlled its release into the surrounding tissue. The filling solution was a water-propanol mix containing 20% L-arginine. The coating solution was comprised of poly-ethylene-glycol (PEG-8000) and dexamethasone. The filling quality was investigated after altering the following variables: plasma surface activation (type of gas, pressure, power, and duration), water-propanol percentage ratio of the filling solution, lifting speed from the bath, and effect of ultrasonic vibration (monofrequency versus multifrequency). The surface coating was evaluated by altering the PEG-8000-dexamethasone percentage ratio and recording the effects on coating thickness and structure, on elution rate, and on wear resistance. The optimized process is presented in detail.
Keywords
This publication has 29 references indexed in Scilit:
- Novel tubulin-targeting agents: anticancer activity and pharmacologic profile of epothilones and related analoguesAnnals of Oncology, 2007
- Sirolimus-eluting stent or paclitaxel-eluting stent vs balloon angioplasty for prevention of recurrences in patients with coronary in-stent restenosis: A randomized controlled trialACC Current Journal Review, 2005
- Intravascular Stents in the Last and the Next 10 YearsJournal of Endovascular Therapy, 2004
- Coronary in-stent restenosis: Current status and future strategiesJACC, 2002
- Stent-Based Delivery of Sirolimus Reduces Neointimal Formation in a Porcine Coronary ModelCirculation, 2001
- Rapamycin inhibits vascular smooth muscle cell migration.Journal of Clinical Investigation, 1996
- A Comparison of Balloon-Expandable-Stent Implantation with Balloon Angioplasty in Patients with Coronary Artery DiseaseNew England Journal of Medicine, 1994
- A Randomized Comparison of Coronary-Stent Placement and Balloon Angioplasty in the Treatment of Coronary Artery DiseaseNew England Journal of Medicine, 1994
- Prospects for site-specific delivery of pharmacologic and molecular therapiesJACC, 1994
- Balloon-expandable intracoronary stents in the adult dog.Circulation, 1987