TTF-Modified Biosensors for Hydrogen Peroxide
- 1 June 1994
- journal article
- research article
- Published by Taylor & Francis in Analytical Letters
- Vol. 27 (8) , 1443-1452
Abstract
Amperometric biosensors for hydrogen peroxide were produced, based on TTF-modified graphite disc electrodes, incorporating immobilised peroxidase. At the optimum operating potential (−200 mV vs. calomel electrode), a linear response to hydrogen peroxide was obtained in the range 0–2 mM (Km value = 1.98 mM). Low sensitivity to pH fluctuations and good stability at room temperature were exhibited.Keywords
This publication has 7 references indexed in Scilit:
- Organic-phase biosensors for monitoring phenol and hydrogen peroxide in pharmaceutical antibacterial productsThe Analyst, 1993
- Lettuce seed meal tissue‐based membrane electrode with high biocatalytic activity for hydrogen peroxideElectroanalysis, 1992
- The polycrystalline fluoride ion-selective electrode and horseradish peroxidase—an alternative electrochemical biosensor?Biosensors and Bioelectronics, 1992
- Pulsed amperometric detection of glucose using a mediated enzyme electrodeJournal of Electroanalytical Chemistry and Interfacial Electrochemistry, 1990
- Biosensor for determination of hydrogen peroxide based on catalase activity of human erythrocytesAnalytica Chimica Acta, 1990
- Amperometric tetrathiafulvalene-mediated lactate electrode using lactate oxidase absorbed on carbon foilAnalytica Chimica Acta, 1990
- Ferrocene-mediated enzyme electrode for amperometric determination of glucoseAnalytical Chemistry, 1984