Simulation of 4H-SiC detectors for ultra fast particle spectroscopy
Authors:
- 1 March 2015
- journal article
- Published by IOP Publishing in Journal of Instrumentation
- Vol. 10 (03) , P03013
Abstract
We present a study on the high speed detection capability of Nickel/4H-SiC Schottky junctions for protons and alpha particles with different energies in the MeV range. The particle path and the distribution of the deposited energy were calculated by SRIM. The detector current signals are calculated by adding the contributions of all generated charge along the particle path. The voltage signals across a 50 Ω resistor load are also derived considering the detector capacitance. Simulations have been performed for different particle energies and detector bias voltages. The fastest simulated response to a 5.5 MeV alpha particle shows a rise-time of 330 ps and a pulse width of 730 ps FWHM, which are in good agreement with experimental values. Similar values have been predicted for 2 MeV protons.Keywords
This publication has 6 references indexed in Scilit:
- Study of the time response of 4H-SiC Schottky junctions for radiation high speed detectionJournal of Instrumentation, 2014
- Characterizing the Timing Performance of a Fast 4H-SiC Detector With an $^{241}$ Am SourceIEEE Transactions on Nuclear Science, 2013
- Silicon carbide detector for laser-generated plasma radiationApplied Surface Science, 2013
- Radiation Detection Properties of 4H-SiC Schottky Diodes Irradiated Up to$10^16$n/cm$^2$by 1 MeV NeutronsIEEE Transactions on Nuclear Science, 2006
- Average energy dissipated by mega-electron-volt hydrogen and helium ions per electron-hole pair generation in 4H-SiCApplied Physics Letters, 2005
- SiC X-Ray Detectors for Spectroscopy and Imaging over a Wide Temperature RangeMaterials Science Forum, 2003