Handheld Fluorometers for Lab-on-a-Chip Applications

Authors:
  • NNNicole Nelson
    Nicole Nelson
  • DSDavid Sander
    David Sander
  • SPSomashekar Bangalore Prakash
    Somashekar Bangalore Prakash
  • ASAnshu Sarje
    Anshu Sarje
  • PAPamela Abshire
    Pamela Abshire

Abstract
We describe the design, fabrication, and performance of a class of simple handheld fluorometers. The devices consist of a sensor along with an integrated optical filter packaged in a handheld format. The sensor is a differential active pixel sensor with in-pixel correlated double sampling fabricated in a 0.5-mu m 2-poly 3-metal complementary metal-oxide semiconductor process and has a readout noise of 175.3 muV, reset noise of 360 muV, dynamic range of 59 dB, and conversion gain of 530 nV/e- . The filter is a high rejection chromophore embedded in a polymer film which is cast onto the chip. We show the results of bioassays utilizing two different single color fluorometers constructed by using the chromophores 2-(2'-hydroxy 5'-methylphenyl) benzotriazole and Sudan II with long-pass wavelengths of 400 nm and 540 nm, respectively. The bioassays measures metabolic activity and viability of biological cells, which are useful for cytotoxicity and pathogen detection applications.

This publication has 33 references indexed in Scilit: