Design and evaluation of microheater combined QCM array for thermal desorption spectroscopy

Authors:

Abstract
In this paper, we propose a single microheater combined quartz crystal microbalance array for thermal desorption spectroscopy. The each QCM has two electrodes for exciting thickness shear mode (TSM) vibrations on one side of the crystal and a double-spiral-shape heater on the other side. The resonance frequency is stable when a voltage was applied on the microheater. Carbon microparticles for gas adsorption were coated on one of the QCMs and the frequency change during temperature increase was extracted by subtracting the change of the other QCM. The proposed QCM array can monitor small mass changes at sub-ng level as a function of temperature.