Design and evaluation of microheater combined QCM array for thermal desorption spectroscopy
Authors:
- 1 June 2013
- conference paper
- Published by Institute of Electrical and Electronics Engineers (IEEE)
- No. 2159547X,p. 234-237
Abstract
In this paper, we propose a single microheater combined quartz crystal microbalance array for thermal desorption spectroscopy. The each QCM has two electrodes for exciting thickness shear mode (TSM) vibrations on one side of the crystal and a double-spiral-shape heater on the other side. The resonance frequency is stable when a voltage was applied on the microheater. Carbon microparticles for gas adsorption were coated on one of the QCMs and the frequency change during temperature increase was extracted by subtracting the change of the other QCM. The proposed QCM array can monitor small mass changes at sub-ng level as a function of temperature.Keywords
This publication has 6 references indexed in Scilit:
- A review of monolithic multichannel quartz crystal microbalance: A reviewAnalytica Chimica Acta, 2011
- Quartz Crystal Microbalances for Microscale Thermogravimetric AnalysisAnalytical Chemistry, 2010
- Real-time measurement of photocatalytic reactions using a monolithic QCM arrayJournal of Micromechanics and Microengineering, 2009
- Design and Evaluation of an Antiparallel Coupled Resonator for Chemical Sensor ApplicationsAnalytical Chemistry, 2007
- A gas chromatography quartz crystal microbalance for speciation of nitroaromatic compounds in landfill gasTalanta, 2001
- Discrimination of odorants of definite concentrations by using plasma-organic-film-coated QCR sensorsSensors and Actuators B: Chemical, 2000