Comparability of Pulmonary Nodule Size Measurements among Different Scanners and Protocols: Should Diameter Be Favorized over Volume?
Open Access
- 8 February 2023
- journal article
- research article
- Published by MDPI AG in Diagnostics
- Vol. 13 (4) , 631
Abstract
Background: To assess the impact of the lung cancer screening protocol recommended by the European Society of Thoracic Imaging (ESTI) on nodule diameter, volume, and density throughout different computed tomography (CT) scanners. Methods: An anthropomorphic chest phantom containing fourteen different-sized (range 3–12 mm) and CT-attenuated (100 HU, −630 HU and −800 HU, termed as solid, GG1 and GG2) pulmonary nodules was imaged on five CT scanners with institute-specific standard protocols (PS) and the lung cancer screening protocol recommended by ESTI (ESTI protocol, PE). Images were reconstructed with filtered back projection (FBP) and iterative reconstruction (REC). Image noise, nodule density and size (diameter/volume) were measured. Absolute percentage errors (APEs) of measurements were calculated. Results: Using PE, dosage variance between different scanners tended to decrease compared to PS, and the mean differences were statistically insignificant (p = 0.48). PS and PE(REC) showed significantly less image noise than PE(FBP) (p < 0.001). The smallest size measurement errors were noted with volumetric measurements in PE(REC) and highest with diametric measurements in PE(FBP). Volume performed better than diameter measurements in solid and GG1 nodules (p < 0.001). However, in GG2 nodules, this could not be observed (p = 0.20). Regarding nodule density, REC values were more consistent throughout different scanners and protocols. Conclusion: Considering radiation dose, image noise, nodule size, and density measurements, we fully endorse the ESTI screening protocol including the use of REC. For size measurements, volume should be preferred over diameter.Keywords
Funding Information
- Iten-Kohaut Foundation, Switzerland
- CRPP “AI Oncological Imaging Network of the University of Zurich”
This publication has 30 references indexed in Scilit:
- Ultralow-Dose Chest Computed Tomography for Pulmonary Nodule DetectionInvestigative Radiology, 2014
- Computed tomography attenuation predicts the growth of pure ground-glass nodulesLung Cancer, 2014
- Iterative reconstruction technique vs filter back projection: utility for quantitative bronchial assessment on low-dose thin-section MDCT in patients with/without chronic obstructive pulmonary diseaseEuropean Radiology, 2014
- Semi-Automatic Quantification of Subsolid Pulmonary Nodules: Comparison with Manual MeasurementsPLOS ONE, 2013
- Management strategy of pulmonary nodule in 2013Diagnostic and Interventional Imaging, 2013
- Reduced Lung-Cancer Mortality with Low-Dose Computed Tomographic ScreeningNew England Journal of Medicine, 2011
- A comparison of six software packages for evaluation of solid lung nodules using semi-automated volumetry: What is the minimum increase in size to detect growth in repeated CT examinationsEuropean Radiology, 2008
- Principles of CT: Radiation Dose and Image QualityJournal of Nuclear Medicine Technology, 2007
- Pulmonary Nodules Detected at Lung Cancer Screening: Interobserver Variability of Semiautomated Volume MeasurementsRadiology, 2006
- Interobserver and Intraobserver Variability in the Assessment of Pulmonary Nodule Size on CT Using Film and Computer Display Methods1Academic Radiology, 2005