Simultaneous time- and wavelength-resolved fluorescence spectroscopy for near real-time tissue diagnosis
- 14 March 2008
- journal article
- Published by Optica Publishing Group in Optics Letters
- Vol. 33 (6) , 630-632
Abstract
A novel fiber-optic-based method for simultaneous time- and wavelength-resolved fluorescence spectroscopy for the rapid diagnosis of diseased tissue is demonstrated. By combining multiple bandpass and dichroic filters (405/40, 460/50, and 550/50) with different lengths of optical fiber (1, 10, and ) acting as an optical delay this system enables the near real-time acquisition and characterization of time-resolved fluorescence spectra using a single detector and excitation input. The recording of multiple fluorescence response pulses at selected wavelengths can be completed in hundreds of nanoseconds, which provides the capability of a real-time characterization of biological systems.
Keywords
This publication has 5 references indexed in Scilit:
- High-speed wide-field time-gated endoscopic fluorescence-lifetime imagingOptics Letters, 2004
- Intramolecular Fluorescence Quenching of Crowned 7-Aminocoumarins as Potential Fluorescent ChemosensorsJournal of Fluorescence, 2004
- Time-domain laser-induced fluorescence spectroscopy apparatus for clinical diagnosticsReview of Scientific Instruments, 2003
- Studying biological tissue with fluorescence lifetime imaging: microscopy, endoscopy, and complex decay profilesApplied Optics, 2003
- Temporally and spectrally resolved fluorescence spectroscopy for the detection of high grade dysplasia in Barrett's esophagusLasers in Surgery and Medicine, 2003