Integration of uniform porous shell layers in very long pillar array columns using electrochemical anodization for liquid chromatography
- 31 October 2013
- journal article
- research article
- Published by Royal Society of Chemistry (RSC) in The Analyst
- Vol. 139 (3) , 618-625
Abstract
Electrochemical anodization has been applied to grow porous shell layers of 300 nm (30 nm pores) in 5 μm diameter pillar array columns (PACs) with a spacing of 2.5 μm. Using turn structures preceded and followed by the flow distributor structures recently introduced by our group and filled with radially elongated pillars, columns with quasi unlimited channel lengths could be conceived. The uniformity of the porous PAC was assessed by determining local plate heights along the channel, which appeared to be constant. Minimal (absolute) plate heights (H) between 4 and 6 μm were obtained at optimal flow rates when imposing increasing retention factors. Upon measuring the surface area involved in chromatographic retention as an indicator of the available surface area, an increase in the surface area by a factor of about 30 compared to that of non-anodized pillars was found. On reconfiguring a commercial HPLC instrument to enable on-chip injections, 90% of the performance (expressed in theoretical plates) could be maintained for a 1 m column, while for a 25 cm column severe losses were still observed. As the corresponding pressure drop for optimal operation of retained components is on the order of 10 bar per m only, portable and cheaper HPLC devices with high efficiencies become realistically conceivable.Keywords
This publication has 28 references indexed in Scilit:
- Capillary liquid chromatography separations using non-porous pillar array columnsJournal of Chromatography A, 2012
- Estimation and optimization of the peak capacity of one-dimensional gradient high performance liquid chromatography using a long monolithic silica capillary columnJournal of Chromatography A, 2012
- Quest for organic polymer-based monolithic columns affording enhanced efficiency in high performance liquid chromatography separations of small molecules in isocratic modeJournal of Chromatography A, 2012
- Realization of 1 × 106 Theoretical Plates in Liquid Chromatography Using Very Long Pillar Array ColumnsAnalytical Chemistry, 2012
- High-Aspect-Ratio, Silicon Oxide-Enclosed Pillar Structures in Microfluidic Liquid ChromatographyAnalytical Chemistry, 2010
- Use of Folded Micromachined Pillar Array Column with Low-Dispersion Turns for Pressure-Driven Liquid ChromatographyAnalytical Chemistry, 2010
- Comparison study of porous, fused-core, and monolithic silica-based C18 HPLC columns for Celestoderm-V Ointment® analysisJournal of Pharmaceutical and Biomedical Analysis, 2009
- Design and evaluation of flow distributors for microfabricated pillar array columnsLab on a Chip, 2009
- Experimental Study of Porous Silicon Shell Pillars under Retentive ConditionsAnalytical Chemistry, 2008
- Experimental Study on Band Dispersion in Channels Structured with MicropillarsAnalytical Chemistry, 2006