Suitability of Different Blood-Analogous Fluids in Determining the Pump Characteristics of a Ventricular Assist Device
Open Access
Abstract
Ventricular assist devices (VADs) are implantable turbomachines that save and improve the lives of patients with severe heart failure. In the preclinical evaluation, a VAD design must be experimentally or numerically tested regarding its pump characteristics, primarily for its pressure buildup (pressure head ) since it must provide the cardiovascular system with a sufficient blood flow rate . Those pump characteristics are determined on a test bench. Here, a glycerol-water mixture is almost exclusively used as blood-analogous fluid, which should reflect the properties (density, viscosity) of blood as close as possible. However, glycerol water has some disadvantages, such as a higher density compared to real blood and a relatively high cost. Therefore, the study aimed to analyze six different blood analogous fluids to select the most suitable one in consideration of fluid handling, costs, and, most importantly, fluid properties (material and rheological). First, all fluids were mixed to achieve reference values of blood density and viscosity from the literature. Afterwards, the pump characteristics (pressure heads and efficiencies via the VAD) were experimentally and numerically determined and compared among each other and with literature values. Of all six investigated fluids, only the aqueous–polyethylene glycol 200 (PEG 200) solution matches exactly the desired blood properties, and the pump characteristics of this fluid are in the expected range for the analyzed operation point of the VAD. Another advantage is that the cost of the mixture is 35% lower compared to glycerol water. Additionally, we demonstrate that non-Newtonian flow behavior has little effect on the pump characteristics in our VAD.
Keywords
Funding Information
- German Research Foundation (469384587)
This publication has 28 references indexed in Scilit:
- Mock Circulation Loop to Investigate Hemolysis in a Pulsatile Total Artificial HeartArtificial Organs, 2015
- Dynamic viscosity versus probe-reported microviscosity of aqueous mixtures of poly(ethylene glycol)The Journal of Chemical Thermodynamics, 2012
- Validation of an Axial Flow Blood Pump: Computational Fluid Dynamics Results Using Particle Image VelocimetryArtificial Organs, 2011
- Numerical Design and Experimental Hydraulic Testing of an Axial Flow Ventricular Assist Device for Infants and ChildrenASAIO Journal, 2007
- Experimental Study on Shear Stress Distributions in a Centrifugal Blood PumpArtificial Organs, 2007
- Matching index of refraction using a diethyl phthalate/ethanol solution for in vitro cardiovascular modelsExperiments in Fluids, 2006
- Geometrical Aspects During Formation of Compact Aggregates of Red Blood CellsMaterials Research, 2002
- Investigation of fluid dynamics within a miniature mixed flow blood pumpExperiments in Fluids, 2001
- The physical properties of blood – forensic considerationsScience & Justice, 1996
- Density and Viscosity of Concentrated Aqueous Solutions of Polyethylene GlycolJournal of Chemical & Engineering Data, 1994