Feasibility of capturing vessel expansion with 4D‐CTA: Phantom study to determine reproducibility, spatial and temporal resolution
Authors:
Open Access
- 12 August 2024
- journal article
- research article
- Published by Wiley in Medical Physics
- Vol. 51 (10) , 7171-7179
Abstract
Background: Dynamic Computed Tomography Angiography (4D CTA) has the potential of providing insight into the biomechanical properties of the vessel wall, by capturing motion of the vessel wall. For vascular pathologies, like intracranial aneurysms, this could potentially refine diagnosis, prognosis, and treatment decision‐making.Purpose: The objective of this research is to determine the feasibility of a 4D CTA scanner for accurately measuring harmonic diameter changes in an in‐vitro simulated vessel.Methods: A silicon tube was exposed to a simulated heartbeat. Simulated heart rates between 40 and 100 beats‐per‐minute (bpm) were tested and the flow amplitude was varied, resulting in various changes of tube diameter. A 320‐detector row CT system with ECG‐gating captured three consecutive cycles of expansion. Image registration was used to calculate the diameter change. A vascular echography set‐up was used as a reference, using a 9 MHz linear array transducer. The reproducibility of 4D CTA was represented by the Pearson correlation (r) between the three consecutive diameter change patterns, captured by 4D CTA. The peak value similarity (pvs) was calculated between the 4D CTA and US measurements for increasing frequencies and was chosen as a measure of temporal resolution. Spatial resolution was represented by the Sum of the Relative Percentual Difference (SRPD) between 4D CTA and US diameter change patterns for increasing amplitudes.Results: The reproducibility of 4D CTA measurements was good (r ≥ 0.9) if the diameter change was larger than 0.3 mm, moderate (0.7 ≤ r < 0.9) if the diameter change was between 0.1 and 0.3 mm, and low (r < 0.7) if the diameter change was smaller than 0.1 mm. Regarding the temporal resolution, the amplitude of 4D CTA was similar to the US measurements (pvs ≥ 90%) for the frequencies of 40 and 50 bpm. Frequencies between 60 and 80 bpm result in a moderate similarity (70% ≤ pvs < 90%). A low similarity (pvs < 70%) is observed for 90 and 100 bpm. Regarding the spatial resolution, diameter changes above 0.30 mm result in SRPDs consistently below 50%.Conclusion: In a phantom setting, 4D CTA can be used to reliably capture reproducible tube diameter changes exceeding 0.30 mm. Low pulsation frequencies (40 or 50 bpm) provide an accurate measurement of the maximum tube diameter change.Keywords
This publication has 15 references indexed in Scilit:
- Photon-counting CT: Review of initial clinical resultsEuropean Journal of Radiology, 2023
- Development of experimental ground truth and quantification of intracranial aneurysm pulsation in a patientScientific Reports, 2021
- Dynamic Volume Change Rate and Aspect Ratio Are Correlated to the Formation of an Irregular Morphology of Unruptured Intracranial AneurysmJournal of Computer Assisted Tomography, 2019
- Biomechanical Characterization of Intracranial Aneurysm Wall: A Multiscale StudyWorld Neurosurgery, 2018
- Detection of Pulsation in Unruptured Cerebral Aneurysms by ECG-Gated 3D-CT Angiography (4D-CTA) with 320-Row Area Detector CT (ADCT) and Follow-up Evaluation Results: Assessment Based on Heart Rate at the Time of ScanningClinical Neuroradiology, 2013
- Quantification of Intracranial Aneurysm Morphodynamics from ECG-gated CT AngiographyAcademic Radiology, 2013
- Cardiac Cycle-Related Volume Change in Unruptured Cerebral AneurysmsStroke, 2012
- Performance Evaluation of Methods for Two-Dimensional Displacement and Strain Estimation Using Ultrasound Radio Frequency DataUltrasound in Medicine & Biology, 2009
- Statistical Properties of Jacobian Maps and the Realization of Unbiased Large-Deformation Nonlinear Image RegistrationIEEE Transactions on Medical Imaging, 2007
- Intensity Gradient Based Registration and Fusion of Multi-modal ImagesMethods of Information in Medicine, 2007