Long-Term Clinical and Echocardiographic Follow-Up of the Freestyle Stentless Aortic Bioprosthesis
- 11 September 2012
- journal article
- research article
- Published by Wolters Kluwer Health in Circulation
- Vol. 126 (11_suppl_1) , S198-204
Abstract
Background—: Stentless aortic bioprostheses were designed to provide enhanced hemodynamic performance and potentially greater longevity. The present report describes the outcomes of patients with the Freestyle stentless bioprosthesis followed for ≤18 years. Methods and Results—: Between 1993 and 2011, 430 patients underwent primary aortic valve replacement with a Freestyle bioprosthesis in the subcoronary position. Mean age was 68.2±8.2 years. All of the clinical and echocardiographic data were collected prospectively. Mean overall follow-up was 9.1±4.4 years and was complete in all of the patients. In-hospital mortality was 3.5% (n=15). Overall, 10- and 15-year survival were 60.7% and 35.0%, respectively. Fifty-one patients required reoperation during follow-up, including 27 for structural valve deterioration (SVD). Overall, freedom from reoperation was 91.0% and 75.0% at 10 and 15 years, whereas freedom from reoperation for SVD was 95.9% and 82.3%, respectively. At 10 and 15 years, freedom from reoperation for SVD was 94.0% and 62.6% for patients P =0.002). The median time to explant for SVD was 10.7 years. SVD presented mostly as acute, severe aortic insufficiency attributed to leaflet tear (77.8%). The independent risk factors for reoperation for SVD were age 60 years of age. Severe aortic insufficiency with leaflet tear is the major mode of SVD leading to reoperation in these patients.Keywords
This publication has 33 references indexed in Scilit:
- Structural deterioration of the Freestyle aortic valve: Mode of presentation and mechanismsThe Journal of Thoracic and Cardiovascular Surgery, 2006
- Ten-Year Outcome After Aortic Valve Replacement with the Freestyle Stentless BioprosthesisThe Annals of Thoracic Surgery, 2005
- Eight-year results of aortic root replacement with the freestyle stentless porcine aortic root bioprosthesisThe Annals of Thoracic Surgery, 2002
- Are stentless valves hemodynamically superior to stented valves? A prospective randomized trialThe Annals of Thoracic Surgery, 2002
- Comparative rest and exercise hemodynamics of 23-mm stentless versus 23-mm stented aortic bioprosthesesThe Annals of Thoracic Surgery, 2000
- Hemodynamic and physical performance during maximal exercise in patients with an aortic bioprosthetic valveJACC, 1999
- Hemodynamic features of the freestyle aortic bioprosthesis compared with stented bioprosthesisThe Annals of Thoracic Surgery, 1998
- Medtronic Freestyle Aortic Root Bioprosthesis: Implant TechniquesJournal of Cardiac Surgery, 1998
- Guidelines for reporting morbidity and mortality after cardiac valvular operationsThe Journal of Thoracic and Cardiovascular Surgery, 1996
- Validation of applications of indexed aortic prosthetic valve areas calculated by Doppler echocardiographyJACC, 1990