Comparison of theEPOCand i‐STATanalysers for canine blood gas and electrolyte analysis

Authors:

Abstract
OBJECTIVES: A prospective study to evaluate agreement and precision of a new point‐of‐care portable analyser, theEPOCanalyser, compared with the i‐STATanalyser in canine blood.METHODS: Blood samples (68 venous and 32 arterial) were obtained from 63 client‐owned dogs for clinical reasons and surplus blood was used to analyse agreement between theEPOCand i‐STATanalysers. Precision of theEPOCanalyser was also assessed by repeat analysis of 20 samples. Measured analytes includedpH, partial pressures of carbon dioxide and oxygen and concentrations of sodium, potassium, ionised calcium, glucose and haematocrit. Haemoglobin, base excess, bicarbonate and saturation of haemoglobin with oxygen were calculated.RESULTS: EPOCprecision was acceptable, but agreement was poor for sodium, haematocrit, haemoglobin and base excess. Overall, method comparison was poor forpH, partial pressure of oxygen, sodium, haematocrit, haemoglobin and base excess.CLINICAL SIGNIFICANCE: TheEPOCanalyser is useful for dogs, although clinically significant differences between theEPOCand i‐STATanalysers exist for some analytes, and as such these analysers should not be used interchangeably.