Thorough preoperative planning in total knee arthroplasty is essential to reduce implant failure by proper implant sizing and alignment. The “gold standard” in conventional preoperative planning is based on anterior-posterior long-leg radiographs. However, the coronal component alignment is still an open discussion in literature, since studies have reported contradictory outcomes on survivorship, indicating that optimal individual alignment goals still need to be defined. Two-dimensional biomechanical models of the knee have the potential to predict joint forces and, therefore, objectify therapy planning. Previously published two-dimensional biomechanical models were evaluated and validated for the first time in this study by comparison of model predictions to corresponding in vivo measurements obtained from telemetric implants for a one- and two-leg stance. Model input parameters were acquired from weight-bearing anterior-posterior long-leg radiographs and statistical assumptions for patient-specific model adaptation. The overall time from initialization to load prediction was in the range of 7–8 minutes per patient for all models. However, no model could accurately predict the correct trend of knee joint forces over patients. Two dimensional biomechanical models of the knee have the potential to improve preoperative planning in total knee arthroplasty by providing additional individual biomechanical information to the surgeon. Although integration into the clinical workflow might be performed with acceptable costs, the models’ accuracy is insufficient for the moment. Future work is needed for model optimization and more sophisticated modelling approaches.Keywords
This publication has 42 references indexed in Scilit:
- The relationship between the survival of total knee arthroplasty and postoperative coronal, sagittal and rotational alignment of knee prosthesisInternational Orthopaedics, 2013
- Why Are Total Knees Failing Today? Etiology of Total Knee Revision in 2010 and 2011The Journal of Arthroplasty, 2013
- The Chitranjan Ranawat Award: Is Neutral Mechanical Alignment Normal for All Patients?: The Concept of Constitutional VarusClinical Orthopaedics and Related Research, 2012
- The effect of post-operative mechanical axis alignment on the survival of primary total knee replacements after a follow-up of 15 yearsThe Journal of Bone and Joint Surgery. British volume, 2011
- The influence of postoperative coronal alignment on revision surgery in total knee arthroplastyInternational Orthopaedics, 2007
- The movement of the normal tibio-femoral jointJournal of Biomechanics, 2005
- The Chetranjan Ranawat Award: Tibial Component Failure Mechanisms in Total Knee ArthroplastyClinical Orthopaedics and Related Research, 2004
- The orientation of the distal part of the quadriceps femoris muscle as a function of the knee flexion-extension angleJournal of Biomechanics, 1985
- Rorces at the knee joint: Anatomical considerationsJournal of Biomechanics, 1981
- A Method for Quantitative Analysis of Medial and Lateral Compression Forces at the Knee During StandingClinical Orthopaedics and Related Research, 1972